Chapter 5 To Trigger the Oscilloscope
Data Type:
Click Data Type continuously to set the effective pattern of the data signal. It can
be set to H (high level) or L (low level).
Trigger Condition:
Click When to select the trigger condition of Setup/Hold trigger.
Setup: the oscilloscope triggers when the setup time is smaller than the
specified setup time. After selecting this type, click More → Setup to set the
setup time. You can use the pop-up numeric keypad or scroll with the mouse to
set it.
Hold: the oscilloscope triggers when the hold time is smaller than the specified
hold time. After selecting this type, click More → Hold to set the hold time. You
can use the pop-up numeric keypad or scroll with the mouse to set it.
Setup/Hold: the oscilloscope triggers when the setup time or the hold time is
smaller than the specified time value. After selecting this type, click More →
Setup and More → Hold to set the setup time and hold time respectively. You
can use the pop-up numeric keypad or scroll with the mouse to set them.
Trigger Mode:
For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameter (noise rejection) under this trigger type. For details, refer
to descriptions in "Noise Rejection".
Trigger Level:
Refer to "Trigger Level". The current trigger level value is displayed at the
upper-right corner of the screen.
DS8000-R User Guide
RIGOL
5-29