Table of Contents

Advertisement

Reference Manual
Temperature Calibrator
JOFRA ATC-125/140/156/157/250/320/650 A/B
 Copyright 2009 AMETEK Denmark A/S

Advertisement

Table of Contents
loading

Summary of Contents for Ametek JOFRA ATC-125 A

  • Page 1 Reference Manual Temperature Calibrator JOFRA ATC-125/140/156/157/250/320/650 A/B  Copyright 2009 AMETEK Denmark A/S...
  • Page 2: About This Manual

    The structure of the manual This reference manual is aimed at users who are familiar with AMETEK calibrators, as well as those who are not. The manual is divided into 9 chapters. These describe how to set up, operate, service and maintain the calibrator. The technical specifications are described and accessories may be ordered from the list of accessories.
  • Page 3: Table Of Contents

    List of contents Introduction ................5 Safety instructions ..............7 Setting up the calibrator ............13 3.1 Receipt of the calibrator.............. 13 3.2 Preparing the dry-block calibrator..........16 3.2.1 When setting up the dry-block calibrator, you must… ... 17 3.2.2 Choice of insertion tube..........19 3.2.3 Inserting the sensors ............
  • Page 4 4.8.7 Selecting the stirrer speed (ATC-140/250 A/B only) ..59 4.8.8 Setting the access code ..........59 4.8.9 Resetting the calibrator setup to factory defaults ..60 4.8.10 About the calibrator............61 4.9 Simulation or training ..............62 After use..................63 5.1 Storing and transporting the calibrator ........
  • Page 5: Introduction

    Congratulations on your new AMETEK JOFRA ATC Calibrator! With this AMETEK JOFRA calibrator, you have chosen an extremely effective instrument, which we hope will live up to all your expectations. Over the past many years, we have acquired extensive knowledge of industrial temperature calibration.
  • Page 6  ISO-9001 certified AMETEK Denmark A/S was ISO-9001 certified in September 1994 by Bureau Veritas Certification Denmark.  CE-label Your new calibrator bears the CE label and conforms to the EMC Directive and the Low- voltage Directive.  Technical assistance Please contact the dealer from whom you acquired the instrument if you require technical assistance.
  • Page 7: Safety Instructions

    Safety instructions Read this manual carefully before using the instrument! In order to avoid any personal injuries and/or damage to the instrument all safety instructions and warnings must be observed. Disposal – WEEE Directive These calibrators contain Electrical and Electronic circuits and must be recycled or disposed of properly (in accordance with the WEEE Directive 2002/96/EC).
  • Page 8  The ATC-125 contains the gases R-1270 and R-704 under pressure. The calibrator must under no conditions be stored or operated at ambient temperatures above 60C (140F). Doing so may cause a hazard. About the front panel:  For B versions only, the sockets on the input module must NEVER be connected to voltages exceeding 5V for the TC/RTD sockets and 45V for the mA/V sockets proportional to ground.
  • Page 9  AMETEK Denmark A/S does not take any responsibility, if the well is filled with other fluids than those recommended.  Liquid baths should only be operated by trained personal.  Heat transfer fluids must only be used in calibrators prepared as a liquid bath.
  • Page 10  Over 100°C/212°F If the calibrator has been heated up to temperatures above 100°C/212°F, you must wait until the instrument reaches a temperature below 100°C/212°F before you switch it off. Caution – Cold surface Below 0°C/32°F (applies only to the ATC-125/140/156/ 157 A/B models) ...
  • Page 11 The oil level rises several centimetres when the temperature is rising. To stop overflow switch off the main power and the oil level will decrease.  Carefully wipe off all silicone oil from the sensor under test to avoid spreading of the silicone oil. ...
  • Page 12  The tip of the sensor should rest at the bottom of the sensor basket for optimum results (liquid baths only).  Be careful not to submerge the handle or wire inlet of the sensor-under-test in the fluid, as this might damage the sensor (liquid baths only).
  • Page 13: Setting Up The Calibrator

    Setting up the calibrator Receipt of the calibrator When you receive the instrument…  Carefully unpack and check the calibrator and the accessories.  Check the parts against the list shown below. If any of the parts are missing or damaged, please contact the dealer who sold the calibrator.
  • Page 14  1 RS232 serial cable  1 tool for insertion tube  1 traceable certificate (A versions)  2 traceable certificates (B versions)  1 reference manual Reference Manual  1 user manual User Manual ATC-125/156/157/320/650 A/B only  1 insertion tube (user specified) ...
  • Page 15 ATC-140/250 A/B only (dry-block)  1 dry block kit – metric consisting of : - 1 insertion tube, multi-hole, metric, with 11 bores (incl. reference bore) - 1 pcs. insulation plug for multi-hole, metric (ATC-140 A/B only)  1 dry block kit – inch consisting of : - 1 insertion tube, multi-hole, inch, with 11 bores (incl.
  • Page 16: Preparing The Dry-Block Calibrator

    Preparing the dry-block calibrator The ATC-B-version has a precision reference input. To achieve the high precision, a set of sensor coefficients relating to the specific sensor must be present in the ATC. Before use of the ATC, ensure that the correct coefficients in the ATC are equal to those from the sensors calibration certificate.
  • Page 17: When Setting Up The Dry-Block Calibrator, You Must

    Fig. 1a 3.2.1 When setting up the dry-block calibrator, you must…  Place the calibrator on an even horizontal surface where you intend to use it. Caution…  Do not use the instrument if the internal fan is out of order.
  • Page 18  Check the voltage on the power control switch (on/off switch (230V/115V)). If the voltage of the power control switch differs from the line voltage, you must adjust the setting of the power control switch as follows (see fig. 1a): Open the fuse box lid using a screwdriver.
  • Page 19: Choice Of Insertion Tube

    3.2.2 Choice of insertion tube Caution… Before using new insertion tubes for calibration, the insertion tubes must be heated up to maximum temperature - 250°C (482°F) / 320°C (608°F) / 650°C (1202°F) - for a period of minimum 30 minutes. In order to ensure the best calibration of your sensors please avoid using insertion tubes in the ATC-320 calibrator, which have been used in the ATC-650 calibrator.
  • Page 20: Inserting The Sensors

    3.1.3 Inserting the sensors Before inserting the sensors and switching on the calibrator, please note the following important warning: Warning  Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block calibrator. These fluids may penetrate the calibrator and cause damage or create poisonous fumes.
  • Page 21 Caution…  The well and the insertion tube must be clean before use.  Scratches and other damage to the insertion tubes should be avoided by storing the insertion tubes carefully when not in use.  The insertion tube must never be forced into the well. The well could be damaged as a result, and the insertion tube may get stuck.
  • Page 22: Preparing The Liquid Bath Calibrator (Atc-140/250 Only)

     Liquid baths should only be operated by trained personal.  AMETEK Denmark A/S does not take any responsibility, if the well is filled with other fluids than those recommended.  Heat transfer fluids must only be used in calibrators with an liquid bath.
  • Page 23: When Setting Up The Liquid Bath Calibrator, You Must

    Reference sensor (if available) Fig. 1b 3.3.1 When setting up the liquid bath calibrator, you must…  Place the calibrator on an even horizontal surface where you intend to use it. Place it in a way that will minimize the risk of tilting.
  • Page 24 Caution…  Do not use the instrument if the internal fan is out of order.  The well must be clean before use.  Ensure a free supply of air to the internal fan located at the bottom of the instrument. The area around the calibrator should be free of draught, dirt, flammable substances, etc.
  • Page 25  Silicone oil – Fill the well with oil according to the tables of recommended oil volume. The recommended volumes must be adjusted to the actual job. For oil tables and further oil information – see section 3.3.2. The sensor basket is marked with an optimum fluid level mark (100%).
  • Page 26 Warning Ensure that the sensor is absolutely clean and dry, as a few drops of water might cause a steam explosion.  Place the sensor to be calibrated vertically into the well. It is recommended to use the optional support rod set for a correct position during calibration.
  • Page 27: Selection Of Fluids

    AMETEK Denmark A/S recommends DOW CORNING 200(R) oil 10cSt for sub zero temperatures to 140°C and DOW CORNING 200(R) oil 50cSt for ambient temperature to 250°C. Technical specifications for the liquid baths has been produced using the fluids mentioned above.
  • Page 28: Handling Of Lids

    Tables of recommended oil level @23°C well temperature ATC-250 A/B ATC-140 A/B For recommended For recommended 50 cSt oil 10 cSt oil 0°C - 50°C 100% -20°C - 50°C 100% 50°C - 100°C 50°C - 100°C 100°C - 150°C 100°C - 140°C 150°C - 200°C 200°C - 250°C For the best...
  • Page 29: Inserting The Sensors

    To be able to use the lid for calibration, holes must be drilled in to it, in order to fit your calibration needs. If you use many different sizes of sensors more lids can be purchased at your JOFRA supplier. It is advisable to drill the holes at the same size as the sensors plus 0,5mm and distribute the holes evenly over the lid.
  • Page 30: Operating The Calibrator

    Operating the Calibrator Keyboard, display and standard connections Keyboard CALIBRATION INSTRUMENTS (heading) (icon) (information) (functions) Fig. 3 Pos. Description  LCD.  SOFT KEYS used to select menu options displayed in the LCD.  ENTER KEY used to accept selected options or entered values.
  • Page 31  INFORMATION KEY used to display the status of the parameters involved with the function currently selected.  ESC KEY (escape key) used to cancel a selection/edit or return to previous menu. Display (heading) (icon) (information) (functions) Fig. 4 The Display is divided into four separate areas: ...
  • Page 32 Standard connections (all versions) Fuse 230V 5AF/115V 10AF 115V Sync. RS232 Fig. 5 Pos. Description  Power control switch with a cable connection and on/off switch. It also contains the main fuse. See section 6.1 for information on how to change the fuses and section 3.2.1 to adjust the voltage setting of the power control switch.
  • Page 33: Input Module (B Versions Only)

     Connection for RS232 communication. Input module (B versions only) Warning  The sockets on the input module must NEVER be connected to voltages exceeding 5V for the TC/RTD sockets and 45V for the mA/V sockets proportional to ground.  Thermostats must not be connected to any other voltage sources during test Description of sockets for external connections Switch Test...
  • Page 34: Display Readouts

    Pos. Description  Input for RTD sensor (2, 3 or 4 wire).  Connection to chassis (earth/ground).  Input for reference sensor.  Voltage input.  Passive mA input. Active mA input with 24V supply for transmitter.  Connection for thermostat test. ...
  • Page 35: Main Screen Temperature Values

    The functions are available using the soft keys and are described in sections 4.4 to 4.8. 4.3.1 Main screen temperature values Two temperatures are displayed (A and B versions):  READ temperature: This is the temperature measured by the internal reference sensor. ...
  • Page 36: Set Temperature Menu

    The temperature must be within a range of ±0.03°C/ 0.05°F in 10 minutes to be stable. The criteria can be changed, however, if the temperature stability criteria is set wider or the stability time is set shorter, the calibrator may not reach the SET temperature.
  • Page 37: Running A Calibration

    Note: Calibration information is available in several places throughout the calibration menus. The content of this information is described in section 4.5.3. 4.5.1 Running a calibration  Press to select the Run calibration menu.  to scroll through the list and highlight an existing work order.
  • Page 38  Press to continue the calibration without editing the basic parameters  Press to start the editor. Make the necessary changes, exit the editor by pressing and continue the calibration by pressing If the sensor under test is a thermocouple sensor and the manual compensation mode is selected in work orders, a cold junction temperature must be defined.
  • Page 39 Default value is 0,00C (32F) as if an ice bath is available. Otherwise press to enter another value. Make the necessary changes, exit the editor by pressing continue the calibration by pressing  Press if you wish to overwrite the existing calibration and continue.
  • Page 40: Showing Calibration Results

    The calibration can be stopped at any time, but this will erase all the calibration data. Follow the instructions on screen for repositioning the sensors (if an external manual heat source is used) and entering the step values (if manual input is required). ...
  • Page 41: Displaying Calibration Information

     Press to exit the calibration details and return to the Show calibration results menu. 4.5.3 Displaying calibration information Calibration information is defined within the work orders created on the PC using "JOFRACAL". This information is divided in to six pages of information: ...
  • Page 42: Switch Test Menu

    Scenario (page 1) This shows the calibration setup in a graphic format. The parameters for this setup are defined in the work order created using the PC program. Basic parameters (page 2) This informs you how the calibration was registered, either "as found" or "as left"...
  • Page 43 Hysteresis of a thermostat can also be determined here. Where the hysteresis determines the tolerence between the upper switch temperature and the lower switch temperature of the thermostat. Calibrator stable Heating with Slope Rate Open T [°C/°F] Cooling with Slope Rate Close Open Expected switch range...
  • Page 44: Running A Switch Test

    4.6.1 Running a switch test  Press to select Run switch test. The small graph illustrates the current T and hysteresis selections. Note that T can be greater than T  Press a function soft key ( – ) to enable the editor. –...
  • Page 45: Showing Switch Test Results

    The calibrator's switch test procedure 1. Once the switch test is started, the calibrator starts working towards as quickly as possible. The calibrator's temperature changes (heating or cooling) and switch status are shown in the display. 2. When T is achieved and the temperature is stable, a "" is displayed for one second.
  • Page 46 Results during a switch test  Press to select Show result. This shows the results that are currently available. These results change as the test progresses.  Press to return to the switch test. Finished switch test results At the end of a switch test the results are displayed. These show the final result of the test and are known as the finished results.
  • Page 47: Auto Step Menu

    To view stored finished switch test results  Press to select Show results.  Press a function soft key ( – ) to select the results for one of the last five tests. The data in the information field is the same as that displayed at the end of the switch test.
  • Page 48: Running An Auto Step Calibration

     Press to select the Auto step menu. Two functions are available using the soft keys 4.7.1 Running an Auto step calibration  Press to select Run auto step.  Press to enable the editor to change the Auto step setup and step temperature values.
  • Page 49: Auto Step Test Results

    Make the necessary changes and exit the editor by pressing  Press to start the Auto step test. While the step test is in progress, several functions are available: – To review the Auto step result (no editing is possible). –...
  • Page 50 To view stored switch test results  Press to select Show results.  Press a function soft key ( – ) to select one of the last five Auto step tests stored in the calibrator. 23-03-2009 105446...
  • Page 51: Setup Menu

    Setup menu  Press to select the Setup menu. Nine fuctions are available. These are divided into three separate pages. Use the function key(s) labelled "more" to change page. Use the soft keys ( – ) to select the individual functions: page 1 page 2 page 2...
  • Page 52: Saving A Setup

     Press (setup page 1) to select Load setup.  Use the keyboard to select a calibrator setup number (1–9).  Press to load the selected setup. A warning informs you that the active setup will be overwritten.  Press if you are sure you want to overwrite the existing setup and return to the setup menu.
  • Page 53 Use the function soft keys to set the parameters displayed. Setting the temperature units  Press to select Unit.  Press a function soft key to select the temperature units: - To select Celsius. - To select Fahrenheit. - To select Kelvin ...
  • Page 54 – 0.01° resolution.  Press to accept the new setting and return to the temperature resolution menu. Setting the max. SET temperature  Press to select Max. SET temperature.  Press . A cursor appears in the Max. SET temperature value.
  • Page 55: Setting The Sensor Input Parameters (B Versions Only)

    – Low input temperature that corresponds to the low level electrical signal. – High input (voltage or current). – High input temperature that corresponds to the high level electrical signal.  Use the numeric keys to set a new value or press to edit the existing value.
  • Page 56 Selecting the reference sensor input  Press to select Reference sensor.  Press a function soft key to enable the editor: - To select Internal reference source. Results in displaying the reference as READ. - To select External reference source (reference input on front panel).
  • Page 57 Note that when ON is selected, the calibrator will let the temperature be set by the TRUE temperature. This means it will take longer before the calibrator indicates stable. Note: Set follows TRUE is only relevant when the External reference sensor is displayed in temperature units.
  • Page 58: Altering Stability Criteria

    4.8.6 Altering Stability criteria  Press (setup page 2) to select Stability criteria (ATC-125/ 156/157/320/650 A/B). Press (setup page 2) to select Stability criteria (ATC- 140/250 A/B). The parameters displayed depend on the sensor selected. When none of the parameters displayed are active, then the calibrator's internal reference criteria provide the "time to stable"...
  • Page 59: Selecting The Stirrer Speed (Atc-140/250 A/B Only)

    Selecting the stirrer speed (ATC-140/250 A/B only) 4.8.7  Press (setup page 2) to select Stirrer speed.  Press to edit the speed level of the stirrer. Select a speed setting between 0 and 50. The normal setting is between 10 and 15. When using the ATC-140/250 A/B with a dry block kit the stirrer speed must be set to 0.
  • Page 60: Resetting The Calibrator Setup To Factory Defaults

    The following features can be protected by an access code:  Resetting the calibrator to Factory default settings.  Setting the Maximum SET Temperature.  Editing the Access code while it is enabled.  Press to change the Access code. ...
  • Page 61: About The Calibrator

     Press to restore Default factory settings. 4.8.10 About the calibrator  Press (setup page 3) to select About (ATC-156/157/320/650 A/B).  Press (setup page 3) to select About (ATC-140/250 A/B). This informs you about the calibrator type, the software version installed and the date when it was last calibrated.
  • Page 62: Simulation Or Training

    Simulation or training  Press and hold while you start the calibrator. The calibrator will start in the simulation state. This mode is used to train personnel. The simulation differs from the standard setting in the following ways:  The instrument does not actually heat up or cool down the well.
  • Page 63: After Use

    After use Storing and transporting the calibrator Caution… The following guidelines should always be observed when storing and transporting the calibrator. This will ensure that the instrument and the sensor remain in good working order (all models). Switch off the calibrator using the power control switch. Note that the calibration procedure may be interrupted at any time using the power control switch.
  • Page 64 It is very important that humidity in the well and insertion tube is removed to prevent corrosion and frost expansion damages. Remove the insertion tube from the calibrator using the tool for insertion tube supplied with the instrument (see fig. 9). Tool for insertion tube Hole for tool for insertion tube Insertion tube...
  • Page 65: Transporting The Dry-Block Calibrator

    Caution… The insertion tube must always be removed from the calibrator after use. The humidity in the air may cause verdigris to form on the insertion tube inside the instrument. There is a risk that the insertion tube may become stuck if this is allowed to happen.
  • Page 66: Emptying The Well (Liquidbaths Only)

    Emptying the well (liquid baths only) It is not recommendable to leave the fluid in the well for long-term storage. The best way to store the fluid is in its original airtight container. Warning  Do not handle hot fluid. ...
  • Page 67: Replacing The Main Fuses

    Replacing the main fuses Warning  The calibrator must be switched off before any attempt to service the instrument is made.  The fuse box must not be removed from the power control switch until the mains cable has been disconnected.
  • Page 68: Returning The Calibrator For Service

     parts are replaced/repaired using spare parts which are not identical to those recommended by the manufacturer.  non-original parts are used in any way when operating the instrument. AMETEK Denmark’s liability is restricted to errors that originated from the factory. 23-03-2009 105446...
  • Page 69 ___________________________________________________________________ Safety precautions: if the product has been exposed to any hazardous substances, it must be thoroughly decontaminated before it is returned to AMETEK. Details of the hazardous substances and any precautions to be taken must be enclosed. 105446 06...
  • Page 70: Maintenance

    Maintenance Cleaning Caution… Before cleaning the calibrator, you must switch it off, allow it to cool down and remove all cables. Users should/must carry out the following cleaning procedures as and when required:  The exterior of the instrument - Clean using water or isopropyl alcohol and a soft cloth.
  • Page 71 Warning REMEMBER to wear goggles when using compressed air and cleaning oil. If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form on the well. This, in turn, may cause verdigris to form on the material. To prevent this from happening, the well must be dried. This is done by heating up the calibrator to min.
  • Page 72: Adjusting And Calibrating The Instrument

    Adjusting and calibrating the instrument You are advised to return the calibrator to AMETEK Denmark A/S or an accredited laboratory at least once a year for calibration. Alternatively, you can calibrate/adjust the calibrator yourself using the AmeTrim-ATC Adjust and Calibration Software. This software is divided into 3 separate processes: ...
  • Page 73: Installing The Ametrim-Atc Software

     One vacant RS232 Serial Port (Two are necessary if the DTI is used for the calibration). PC software requirements ® ®  MicrosofWindows 95/98 or Microsoft Windows NT  System font: MS Sans Serif and Arial. 7.2.2 Installing the AmeTrim-ATC Software The software comes on a CD-ROM and is ready to run –...
  • Page 74 Use the normal Windows procedure to start the AmeTrim- ATC software. If you are unsure of how to start software programs, refer to your Windows Help As the software starts, it detects the type of calibrator connected to the PC and reads its serial number. This information plus the COM port the calibrator is connected to is displayed at the bottom of the AmeTrim-ATC window.
  • Page 75: Temperature Adjustment

    Temperature Adjustment This enables you to check the accuracy of the heat source and the internal sensor in the calibrator. This can be done in 3 ways:  Using the external reference sensor on the ATC (B-version) to measure the TRUE temperature, AmeTrim-ATC can perform a fully automatic calibration and adjustment of the internal sensor.
  • Page 76 Sensor Computer RS232 cable Fig. 11  In AmeTrim-ATC main menu, click Temperature adjustment. 23-03-2009 105446...
  • Page 77  Click a radio button to select the scenario that should be used, depending on which calibrated temperature reference that is available.  Click a radio button to select the desired temperature units.  Click a radio button to select whether the calibration only should be performed as an “As Found”...
  • Page 78  Click Yes. The calibrator immediately starts working towards the temperature defined in the first step. A new dialogue containing a table with pre-defined temperature steps is displayed. The calibrator type determines the range of these steps. When using the external reference sensor input on the calibrator, or when using a DTI, the calibration and adjustment will now run automatically.
  • Page 79  When the True temperature is stable click “Enter value”. In the new dialogue, type in the correct “TRUE temperature” value read from the external precision thermometer.  Click OK. The READ temperature measured by the calibrator and the TRUE value are entered in the table. The calibrator now starts working towards the next temperature level.
  • Page 80: Input Adjustment (B Versions Only)

    Note… If you want a record of the measured TRUE and READ values, you must click Print now. These values are not saved once you return to the AmeTrim Main Menu. Note… Downloading new coefficients to the calibrator overwrites the existing ones already resident in the calibrator. To prevent unintentional overwriting of existing coefficients, a warning appears requiring you to confirm the download.
  • Page 81 Nominal values Accepted range  RTD and Reference Inputs: 50Ω (45 to 70Ω) 176Ω (150 to 200Ω) 345Ω (330 to 350Ω)  RTD inputs: 1000Ω (800 to 1200Ω) 2700Ω (2300 to 2900Ω)  mA inputs: 0 mA (-1 to +1 mA) 10 mA (9 to 11 mA) 20 mA...
  • Page 82 Container with ice 0°C reference Principle sketch Fig. 12 If at any point you do not wish to continue, click Stop. The software exits this option and returns to the Main Menu.  In AmeTrim-ATC main menu, click Input adjustment. By clicking the Calibration Date button, you can view the last calibration dates for each of the inputs.
  • Page 83 To view the current Input values  Click the radio button Only measuring.  Use the check box list to select the inputs to be measured. If you have selected more than one input to be measured, the measurements occur in the same order as they appear in the check box list.
  • Page 84  When you are finished viewing or printing the report, click Close. This returns you to the first dialogue in the Temperature adjustment option.  If you do not require any further measurements, click Done to return to the Main Menu. To view, adjust and calibrate the current Input values Note…...
  • Page 85: Reference Sensor

    The software measures the pre-set values for the selected input, adjusts the settings and then re-calibrates the calibrator’s inputs. All this is done before moving on to the next selected input.  When all the selected inputs have been measured, adjusted and re-calibrated a new dialogue appears.
  • Page 86  In AmeTrim-ATC main menu, click Reference sensor. A new dialogue with a table containing the calibration of corresponding temperatures and resistances of the reference sensor. If none are present in the calibrator, then the default values are displayed.  Click a radio button to select the desired temperature units.
  • Page 87: Managing Dti Sensor Coefficients

     If the calibration data fills more than the steps currently displayed, click Add step. If it fills less than the number of steps displayed, click Delete step.  When all the values from the Calibration Certificate have been entered, click Download values. When the values are downloaded they overwrite the ones stored in the calibrator.
  • Page 88: Setup Printer

     Type in the correct coefficients for the selected sensor  When you are satisfied with all the values, click Download. Default values can be selected by using The Default button. 7.2.9 Setup Printer This option provides a standard Windows procedure which enables you to edit the settings for the current printer or change to another printer.
  • Page 89: Technical Specifications

    Technical specifications The illustration below shows the setup that forms the basis for the technical specifications for dry-block calibrators. Fig. 13 105446 06 23-03-2009...
  • Page 90 The illustration below shows the setup that forms the basis for the technical specifications for liquid bath calibrators. Fig. 14 23-03-2009 105446...
  • Page 91 The illustration below shows the setup that forms the basis for the technical specifications for cooling calibrators. Fig. 15 105446 06 23-03-2009...
  • Page 92 Pos. Description – dry-block calibrators (fig. 13)  Calibrator  Ø4.2 mm insertion tube  Insulation plug  Ø4 mm Pt 100 sensor with traceable certificate  DTI-1000 reference precision thermometer with traceable certificate Pos. Description – liquid bath calibrators (fig. 14) ...
  • Page 93 TECHNICAL SPECIFICATIONS – ALL MODELS All specifications are given with an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F GENERAL SPECIFICATIONS MECHANICAL SPECIFICATIONS Dimensions l  w  h 352  156  360 mm (13.9 x 6.1 x 14.2 inch) (ATC-140/156/157/250/320/650) Dimensions l ...
  • Page 94 MECHANICAL SPECIFICATIONS Weight non-drilled insert ATC-140/250 A/B: 1200 g / 42.3 oz ATC-125 / 156 A/B: 290 g / 10.2 oz ATC-157 A/B: 130 g / 4.6 oz ATC-320 / 650 A/B: 940 g / 33.2 oz POWER SUPPLY Line voltage/frequency ATC-125 / 140 / 156 / 157 / 250 / 320 A/B: 90-127VAC / 180-254VAC 45-65 Hz ATC-650 A/B:...
  • Page 95 THERMAL SPECIFICATIONS ATC-125 A/B Well specifications 40 mm / 1.57 inch axial homogeneity: 0.05°C / 0.09°F @125°C / 257°F 0.05°C / 0.09°F @23°C / 73.4°F 0.07°C / 0.13°F @ -90°C / -130°F 50 mm / 1.97 inch axial homogeneity: 0.05°C / 0.09°F @125°C / 257°F 0.05°C / 0.09°F @23°C / 73.4°F 0.10°C / 0.18°F @-90°C / -130°F 60 mm / 2.36 inch axial homogeneity:...
  • Page 96 THERMAL SPECIFICATIONS ATC-125 A/B Long term drift (1 year) : ±0.20°C / ±0.36°F Calibration accuracy (test limit) ±0.13°C / ±0.23°F Temperature coefficient 0.005 °C/°C (0-20°C and 26-40°C) (32-68°F and 78.8-104°F) Stability ±0.03°C / ±0.05°F outside ambient ±5°C / ±9°F ±0.04°C / ±0.07°F ambient ±5°C / ±9°F Reference accuracy ±0.04°C / ±0.07°F Total accuracy...
  • Page 97 THERMAL SPECIFICATIONS – ATC-140 A/B DRY - BLOCK Well specifications 40 mm / 1.57 inch axial homogeneity : 0.05°C / 0.09°F @140°C / 284°F 0.05°C / 0.09°F @-20°C / 4°F 50 mm / 1.97 inch axial homogeneity : 0.10°C / 0.18°F @140°C / 284°F 0.05°C /0.09°F @-20°C / 4°F 60 mm / 2.36 inch axial homogeneity : 0.15°C / 0.27°F @140°C / 284°F...
  • Page 98 THERMAL SPECIFICATIONS – ATC-140 A/B DRY - BLOCK Reference accuracy ±0.02°C / ( ±0.04°F) Total accuracy ±0.18°C / (±0.32°F) Heating time incl. insert -20°C / -4°F to 23°C / 73.4°F : 10 min. 23°C / 73.4°F to 100°C / 212°F: 31 min.
  • Page 99 THERMAL SPECIFICATIONS – ATC-140 A/B LIQUID BATHS Well specifications 40 mm / 1.57 inch axial homogeneity : 0.05°C / 0.09°F @140°C / 284°F 0.05°C / 0.09°F @-20°C / 4°F 50 mm / 1.97 inch axial homogeneity : 0.05°C / 0.09°F @140°C / 284°F 0.05°C /0.09°F @-20°C / 4°F 60 mm / 2.36 inch axial homogeneity : 0.10°C / 0.18°F @140°C / 284°F...
  • Page 100 THERMAL SPECIFICATIONS – ATC-140 A/B LIQUID BATHS Reference accuracy ±0.10°C / ±0.18°F Total accuracy (std. cal. with insert) ±0.30°C / ±0.54°F Time to stability 15 min. THERMAL SPECIFICATIONS ATC-156 A/B Maximum temperature 155°C / 311°F Minimum temperature -40°C / -40°F @ ambient temperature 0°C / 32°F -24°C / -11.2°F @ ambient temperature 23°C / 73.4°F -12°C / -10.4°F @ ambient temperature 40°C /...
  • Page 101 THERMAL SPECIFICATIONS ATC-156 A/B Well specifications 40 mm ( 1.57 inch) axial homogeneity : 0.05°C (0.09°F) @155°C (311°F) 0.05°C ( 0.09°F) @100°C (212°F) 0.05°C ( 0.09°F) @0°C (32°F) 0.05°C ( 0.09°F) @-24°C (-11.2°F) 50 mm (1.97 inch) axial homogeneity: 0.07°C (0.13°F) @155°C (311°F) 0.05°C (0.09°F) @100°C (212°F) 0.06°C (0.11°F) @0°C (32°F) 0.09°C (0.16°F) @-24°C (-11.2°F)
  • Page 102 THERMAL SPECIFICATIONS ATC-156 A/B Influence from load with Ext. Reference : 0.01°C (0.02°F) @155°C (311°F) 0.01°C (0.02°F) @-24°C (-11.2°F) Long term drift (1 year) : ±0.05°C (±0.09°F) Calibration accuracy (test limit) ±0.05°C (0.09°F) Temperature coefficient 5ppm/°C (0-40°C) / (32-104°F) Stability ±0.01°C ( ±0.02F) Reference accuracy ±0.02°C (±0.04°F)
  • Page 103 THERMAL SPECIFICATIONS ATC-157 A/B Maximum temperature 155°C / 311°F Minimum temperature -57°C / -70.6°F @ ambient temperature 0°C / 32°F -45°C / -49°F @ ambient temperature 23°C / 73.4°F -31°C / -23.8°F@ ambient temperature 40°C / 104°F 105446 06 23-03-2009...
  • Page 104 THERMAL SPECIFICATIONS ATC-157 A/B Well specifications 40 mm (1.57 inch) axial homogeneity: 0.05°C (0.09°F) @155°C ( 311°F) 0.05°C (0.09°F) @100°C (212°F) 0.05°C (0.09°F) @ 0°C (32°F) 0.10°C (0.18°F) @-35°C (-31°F) 0.11°C (0.20°F) @-45°C (-49°F) 50 mm (1.97 inch) axial homogeneity: 0.06°C (0.11°F) @155°C (311°F) 0.05°C (0.09°F) @100°C (212°F) 0.07°C (0.13°F) @0°C (32°F)
  • Page 105 THERMAL SPECIFICATIONS ATC-157 A/B Difference between borings : 0.01°C (0.02°F) Influence from load : 0.10°C (0.18°F) @155°C (311°F) 0.12°C (0.22°F) @-45°C (-49°F) Influence from load with Ext. Reference : 0.01°C (0.02°F) @155°C (311°F) 0.01°C (0.02°F) @-45°C ( -49°F) Long term drift (1 year) : ±0.05°C / ±0.09°F Calibration accuracy (test limit) ±0.08°C (±0.14°F)
  • Page 106 THERMAL SPECIFICATIONS – ATC-250 A/B DRY - BLOCK Maximum temperature 250°C / 482°F Minimum temperature 5°C / 41°F @ ambient temperature 0°C / 32°F 28°C / 82°F @ ambient temperature 23°C / 73.4°F 45°C / 113°F@ ambient temperature 40°C / 104°F 23-03-2009 105446...
  • Page 107 THERMAL SPECIFICATIONS – ATC-250 A/B DRY - BLOCK Well specifications 40 mm / 1.57 inch axial homogeneity : 0.10°C / 0.18°F @250°C / 482°F 0.05°C / 0.09°F @140°C / 284°F 0.05°C / 0.09°F @50°C / 122°F 50 mm / 1.97 inch axial homogeneity : 0.20°C / 0.36°F @250°C / 482°F 0.10°C / 0.18°F @140°C / 284°F 0.05°C /0.09°F @50°C / 122°F...
  • Page 108 THERMAL SPECIFICATIONS – ATC-250 A/B DRY - BLOCK Calibration accuracy (test limit) ±0.13°C / ±0.23°F Temperature coefficient 5ppm/°C (0-40°C) (32-104°F) Stability ±0.02°C / ±0.04°F Reference accuracy ±0.06°C (±0.11°F) Total accuracy ±0.28°C (±0.50°F) Heating time incl. insert 28°C / 82°F to 100°C / 212°F : 3 min.
  • Page 109 THERMAL SPECIFICATIONS – ATC-250 A/B LIQUID BATHS Well specifications 40 mm / 1.57 inch axial homogeneity : 0.04°C / 0.07°F @250°C / 482°F 0.04°C / 0.07°F @140°C / 284°F 0.03°C / 0.05°F @50°C / 122°F 50 mm / 1.97 inch axial homogeneity : 0.04°C / 0.07°F @250°C / 482°F 0.04°C / 0.07°F @140°C / 284°F 0.04°C /0.07°F @50°C / 122°F...
  • Page 110 THERMAL SPECIFICATIONS – ATC-250 A/B LIQUID BATHS Long term drift (1 year) : ±0.05°C / ±0.09°F Calibration accuracy (test limit) ±0.13°C / ±0.23°F Temperature coefficient 5ppm/°C (0-40°C) (32-104°F) Stability ±0.02°C / ±0.04°F Reference accuracy ±0.10°C / ±0.18°F Total accuracy (std. cal. with insert) ±0.50°C / ±0.90°F Heating time incl.
  • Page 111 THERMAL SPECIFICATIONS ATC-320 A/B Well specifications 40 mm (1.57 inch) axial homogeneity : 0.15°C (0.27°F) @320°C (608°F) 0.10°C (0.18°F) @155°C (311°F) 0.05°C (0.09°F) @100°C (212°F) 0.05°C (0.09°F) @50°C (122°F) 50 mm (1.97 inch) axial homogeneity : 0.15°C (0.27°F) @320°C (608°F) 0.10°C (0.18°F)@155°C (311°F) 0.05°C (0.09°F) @100°...
  • Page 112 THERMAL SPECIFICATIONS ATC-320 A/B Difference between borings : 0.005°C (0.01°F) @100°C (212°F) 0.01°C (0.02°F) @320°C (608°F) Influence from load : 0.15°C (0.27°F) @320°C (608°F) Influence from load with Ext. Reference : 0.03°C / 0.05°F @320°C / 608°F Long term drift (1 year) : 0.05°C / ±0.09°F Calibration accuracy (test limit) ±0.10°C (±0.18°F)
  • Page 113 THERMAL SPECIFICATIONS ATC-650 A/B Well specifications 40 mm (1.57 inch) axial homogeneity : 0.30°C (0.54°F) @650°C (1202°F) 0.25°C (0.45°F) @320°C (608°F) 0.10°C (0.18°F) @100°C (212°F) 0.05°C (0.09°F) @50°C (122°F) 50 mm (1.97 inch) axial homogeneity : 0.60°C (1.08°F) @650°C (1202°F) 0.35°C (0.63°F) @320°C (608°F) 0.15°C (0.27°F) @100°C (122°F) 0.05°C (0.09°F) @50°C (122°F)
  • Page 114 THERMAL SPECIFICATIONS ATC-650 A/B Influence from load : 0.15°C (0.27°F) @650°C (1202°F) Influence from load with Ext. Reference : 0.10°C (0.18°F) @650°C (1202°F) Long term drift (1 year) : ±0.15°C (±0.27°F) Calibration accuracy (test limit) ±0.10°C (±0.18°F) Temperature coefficient 5ppm/°C (0-40°C) (32-104°F) Stability ±0.01°C (0.02°F) 100°C (212°F) ±0.015°C (0.03°F) 320°C (608°F)
  • Page 115 TECHNICAL SPECIFICATIONS – B MODELS ONLY INPUT SPECIFICATIONS mA input Signal range 0 – 24 mA Internal power supply 24 V, max. 28 mA Resolution 0.001mA / 0.01°C / 0.01°F Accuracy ±(0.010% of rdg. + 0.015% of F.S.) Temperature coefficient 7 ppm/°C (0-40°C) (32-104°F) Input impedance <...
  • Page 116 INPUT SPECIFICATIONS 0.01 / 0.01°C / 0.01°F (Pt500/Pt1000), see page 118 - 119 for accuracy in °C/°F Accuracy (3- or 4-wire) ±(0.005% of rdg. + 0.005% of F.S.) Accuracy (2-wire) ±(0.005% of rdg. + 0.005% of F.S. + 50m) Temperature coefficient 5 ppm/°C (0-40°C) (32-104°F) Type of connection 4 mm safety sockets...
  • Page 117 ACCURACY IN C/F INPUT SPECIFICATIONS ±0.10°C(±0.17°F) @ 320°C(608°F) ±0.12°C(±0.22°F) @ 650°C(1202°F) ±0.19°C(±0.34°F) @ 1200°C(2192°F) Accuracy thermocouple type K input ±0.13°C(±0.24°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.11°C(±0.20°F) @ -50°C(-58°F) ±0.10°C(±0.17°F) @ 0°C(32°F) ±0.11°C(±0.20°F) @ 155°C(311°F) ±0.12°C(±0.22°F) @ 320°C(608°F) ±0.16°C(±0.28°F) @ 650°C(1202°F) ±0.28°C(±0.50°F) @ 1372°C(2502°F) Accuracy thermocouple type T input ±0.14°C(±0.25°F) @ -90°C(-130°F)
  • Page 118 ACCURACY IN C/F INPUT SPECIFICATIONS ±0.48°C(±0.86°F) @ 1820°C(3308°F) Accuracy thermocouple type N ±0.20°C(±0.35°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.16°C(±0.29°F) @ -50°C(-58°F) ±0.15°C(±0.27°F) @ 0°C(32°F) ±0.14°C(±0.24°F) @ 155°C(311°F) ±0.14°C(±0.25°F) @ 320°C(608°F) ±0.16°C(±0.28°F) @ 650°C(1202°F) ±0.17°C(±0.31°F) @ 800°C(1472°F) Accuracy thermocouple type XK ±0.09°C(±0.16°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.07°C(±0.13°F) @ -50°C(-58°F)
  • Page 119 ACCURACY IN C/F INPUT SPECIFICATIONS ±0.050°C(±0.090°F) @ 0°C(32°F) ±0.061°C(±0.110°F) @ 155°C(311°F) ±0.071°C(±0.127°F) @ 320°C(608°F) ±0.087°C(±0.156°F) @ 500°C(932°F) Accuracy RTD Pt500 ±0.079°C(±0.142°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.083°C(±0.149°F) @ -50°C(-58°F) ±0.087°C(±0.157°F) @ 0°C(32°F) ±0.100°C(±0.180°F) @ 155°C(311°F) ±0.111°C(±0.200°F) @ 320°C(608°F) ±0.130°C(±0.235°F) @ 500°C(932°F) Accuracy RTD Pt100 ±0.051°C(±0.092°F) @ -90°C(-130°F) (excluding sensor accuracy)
  • Page 120 ACCURACY IN C/F INPUT SPECIFICATIONS Accuracy RTD Cu100 ±0.047°C(±0.085°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.050°C(±0.090°F) @ -50°C(-58°F) ±0.052°C(±0.094°F) @ 0°C(32°F) ±0.060°C(±0.108°F) @ 150°C(302°F) Accuracy RTD Cu50 ±0.087°C(±0.157°F) @ -90°C(-130°F) (excluding sensor accuracy) ±0.090°C(±0.162°F) @ -50°C(-58°F) ±0.093°C(±0.167°F) @ 0°C(32°F) ±0.100°C(±0.180°F) @ 150°C(302°F) Accuracy Pt100 reference input ±0.019°C(±0.034°F) @ -90°C(-130°F) (excluding sensor accuracy)
  • Page 121: List Of Accessories

    List of accessories All parts listed in the list of accessories can be obtained from the factory through our dealers. Please contact your dealer for assistance if you require parts which do not appear on the list. List of accessories Accessories Parts no.
  • Page 122 List of accessories Accessories Parts no. Extension cable for external probe, banana to lemo 122823 Connector, Lemo male for reference input cable (4.3 to 5.1 mm diameter) 60D711 + 60D712 RS – 232 Interface cable 105366 JOFRACAL PC software 124915 AmeTrim-ATC PC software 105816 Mini Jack connector...
  • Page 123 PARTS NO. FOR STANDARD INSERTION TUBES Sensor ATC-125/156 ATC-157 A/B ATC-320/650 ATC-140 A/B ATC-250 A/B size (Aluminium (Aluminium (Aluminium (Aluminium tubes) (Brass tubes) tubes) tubes) tubes) Undrilled 122720 123286 122719 124899 124891 Undrilled 122722 123285 122721 with ref. 1/8” 105677 123279 105676 3/16”...
  • Page 124 PART NO. FOR STANDARD INSERTION TUBES – MULTI-HOLE Descrip- ATC-125 ATC-156 ATC-157 ATC- ATC-140 ATC-250 tion 320/650 (metric) Type 1 126272 122751 123294 122750 124897 124889 Type 2 126273 122753 123295 122752 Type 3 126274 122755 123296 122754 Type 4 126275 122757 122756...

Table of Contents