System Subsystem; Preset; Posetup <Name - Keithley 7001 Instruction Manual

High density switch system
Hide thumbs Also See for 7001:
Table of Contents

Advertisement

:SYSTem Subsystem

5.17
The system subsystem contains miscellaneous commands and are summarized in Table 5-12.
Table 5-12
SYSTem command summary
Command
:SYSTem

:PRESet

:POSetup <name>
:POSetup?
:VERSion?
:ERRor?
Notes:
1. Brackets ([ ]) are used to denote optional character sets. These optional characters do not have to be included in the program message. Do not use
brackets ([ ]) in the program message.
2. Angle brackets (< >) are used to indicate parameter type. Do not use angle brackets (< >) in the program message.
3. Upper case characters indicate the short-form version for each command word.
4. At least one sapce is required between a command word and the parameter.
5.17.1
Short-form
format
Description
Programming
example
5.17.2
Parameters
Short-form
format
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Description
Return to :SYST:PRES defaults.
Select power-on setup: RST, PRESet, SAV0-SAV9.
Query power-on setup.
Query revision level of SCPI standard.
Query Error Queue.
:PRESet
:SYSTem:PRESet
:syst:pres
This command returns the instrument to states optimized for front panel operation. Table 5-5
defines the default conditions for this command.
OUTPUT 707; ":syst:pres"
:POSetup <name>
:SYSTem:POSetup <name>
<name> = RST
Select *RST defaults on power up
= PRESet
Select :SYSTem:PRESet defaults on power up
= SAVX
Select saved defaults on power up
where; X = 0 to 9
:syst:pos <name>
Return to :SYSTem:PRESet defaults
! Return 7001 to default states
Program power-on defaults
IEEE-488 Reference
Reference
5.17.1
5.17.2
5.17.3
5.17.4
5-117

Advertisement

Table of Contents
loading