Rcl  Recall - Keithley 7001 Instruction Manual

High density switch system
Hide thumbs Also See for 7001:
Table of Contents

Advertisement

IEEE-488 Reference
*RCL  recall
5.10.9
To return the Model 7001 to a setup configuration previously stored in memory.
Purpose
*RCL <n>
Format
Parameters
<NRf> = 0 to 9 Memory location
Description
The Model 7001 can store up to 10 setup configurations in memory locations 0 to 9. Over the
bus, the *SAV command is used to save setup configurations (see paragraph 5.10.11).
The *RCL command is used to return the Model 7001 to a setup configuration stored at one of
the memory locations. For example, if a setup configuration was previously stored at memory
location 9, the Model 7001 can be returned to those setup conditions by sending the following
command:
*RCL 9
Note: The Model 7001 is shipped from the factory with :SYSTem:PRESet defaults stored in all
ten memory locations (see Table 5-5).
Programming
The following example assumes that a setup configuration has been stored at memory location
example
6 (see paragraph 5.10.11 for information on storing setup configurations).
OUTPUT 707; "*RCL 6"
! Recall setup stored at memory location 6.