Power-up
Defaults
*RST
:SYSTem:PRESet
Query
:SCHannel?
command
Short-form formats:
Response message:
The Single Channel mode (when enabled) is used to prevent more than one channel from clos-
Description
ing at a time. When a single channel is closed, the previous closed channel will open. This en-
sures that even with random channel closures, only one channel can be closed at a time.
Toggling the Single Channel mode from off to on causes all channels to open. For complete de-
tails on Single Channel, see Section 4.
Programming
10
example
20
30
40
Line 10 Enables Single Channel and then queries its status.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the status of Single Channel (1; on).
:CPAir <b>
5.13.7
[:ROUTe]:CONFigure:CPAir <b>
Parameters
<b> = 0 or OFF
Short-form
:conf:cpa <b>
format
Power-up
Defaults
*RST
:SYSTem:PRESet
Query
:CPAir?
command
Short-form format:
Response message:
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
[:ROUTe] subsystem
Saved power-on setup
Off
Off
Query status of Single Channel
:conf:sch?
0 (off) or 1 (on)
OUTPUT 707; ":conf:sch on; sch?"
ENTER 707; A$
PRINT A$
END
Disable Card Pair.
= 1 or ON
Enable Card Pair.
Saved power-on setup
Disables Card Pair
Disables Card Pair
Query status of Card Pair
:conf:cpa?
0 (off) or 1 (on)
Control Card Pair