Switching Mainframe B Settings (7001); Run Dut Test System - Keithley 7001 Instruction Manual

High density switch system
Hide thumbs Also See for 7001:
Table of Contents

Advertisement

Front Panel Operation

4.12.3 Switching mainframe B settings (7001)

Scan List = 1!1-1!10, 2!1-2!10, M1,
M1 is a "dummy" channel configured to open all channels. This lets the last channel in a
mainframe open before closing the first channel in the other mainframe.
Card Delay = 0.0
Pole Mode = 2-Pole
Scan Configuration:
Arm Control: Spacing
Input Trigger
Output Trigger
Count
Control
Scan Control:
Channel Control: Spacing
*RESET default setting

4.12.4 Run DUT test SYSTem

1. Press STEP on Switching Mainframe B.
2. Press STEP on Switching Mainframe A.
3. Observe the scanning sequence. The test will stop after two scans of the 40 channels are finished.
4. Press RECALL on the DMM to display the stored readings.
The "Trigger ignored" message will appear on the DMM when the scan sequence switches
from one switching mainframe to the other. This is caused by extraneous triggers and does
not affect the integrity of the testing process. The ignored triggers prevent the "dummy" M1
channels from being measured.
4-66
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
NOTE
TrigLink
Line 4
Line 3
2
Source
Spacing
Input Trigger
Line 4
Output Trigger
Line 3
Count
1
Control
Source
TrigLink
Trigger Link Mode
Asynchronous*
Input Trigger
Line 1*
Output Trigger
Line 2*
Count
ScanList*
Control
Source*
NOTE
TrigLink