Table F-2

Syntax of non-SCPI commands implemented by Model 7001

Command
:DISPlay
:SMESsage <b>
:SMESsage?
:OUTPut
:TTLn
:LSENse AHIGh|ALOW
:LSENse?
[:ROUTe]
:SCAN?
:FCHannels <list>
:FCHannels?
:CONFigure
:SCHannel <b>
:SCHannel?
:CPAir <b>
:CPAir?
:SLOT1
:CTYPE <type>
:CTYPE?
:POLE <n>
:POLE?
:STIMe <num>
:STIMe?
:SLOT2
:CTYPE <type>
:CTYPE?
:POLE <n>
:POLE?
:STIMe <num>
:STIMe?
:MEMory
:SAVe M<num>
:RECall M<num>
:SENSe[1]
:TTL1
:DATA?
:SENSe2
:DATA? <list>
:SENSe3
:DATA? <list>
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Description
Enable (1 or ON) or disable (0 or OFF) displayed status messages.
Request state of status message display.
Path and command to specify output line ÒnÓ (1-4):
Set polarity; active high (AHIG) or active low (ALOW).
Request polarity of speciÞed output line (n).
Request the programmed scan list.
SpeciÞes channels that cannot be closed (forbidden).
Request list of restricted channels.
ConÞguration command path:
Enable (1 or ON) or disable (0 or OFF) Single Channel.
Request status of Single Channel mode.
Enable (1 or ON) or disable (0 or OFF) Channel Pairing.
Request status of Channel Pairing.
Path to conÞgure CARD 1:
Specify the card type installed in the slot.
Request the card type installed in the slot.
Select pole mode (1, 2 or 4).
Request the programmed number of poles.
Set relay settling time in seconds (0 to 99999.999).
Request the programmed relay settling time.
Path to conÞgure CARD 2:
Specify the card type installed in the slot.
Request the card type installed in the slot.
Select pole mode (1, 2 or 4).
Request the programmed number of poles.
Set relay settling time in seconds (0 to 99999.999).
Request the programmed relay settling time.
Path to program memory:
Saves channel pattern at memory location (1-100).
Recalls channel pattern from memory location (1-100).
Command path for input port.
Read digital input port.
Command path:
Read speciÞed input channels for slot 1.
Command path:
Read speciÞed input channels for slot 2.
SCPI Conformance Information