Rst  Reset - Keithley 7001 Instruction Manual

High density switch system
Hide thumbs Also See for 7001:
Table of Contents

Advertisement

IEEE-488 Reference
*RST  reset
5.10.10
Purpose
Format
Description
Programming
example
5-52
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
To reset the Model 7001.
*RST
When the *RST command is sent, the Model 7001 performs the following operations:
1. Returns the Model 7001 to the *RST default conditions (see Table 5-5).
2. Cancels all pending commands.
3. Cancels response to any previously received *OPC and *OPC? commands.
The following instrument states are not affected by this command:
IEEE-488 address
Standard Event Status Register
Status Byte Register
OUTPUT 707; "*RST"
! Return 7001 to *RST defaults