Non-701X Series Cards - Keithley 7001 Instruction Manual

High density switch system
Hide thumbs Also See for 7001:
Table of Contents

Advertisement

Card Installation
CAUTION
To prevent contamination to the
switch card that could degrade
performance, only handle the card
assembly by the edges and shields. If
contamination occurs, clean the card
as explained in its instruction manual.
Card installation
Perform the following steps to install the card assem-
bly in the Model 7001 mainframe:
1. Mate the connector card to the relay card if they are
separate. Make sure to handle the cards by the edg-
es and shields to prevent contamination.
2. Facing the rear panel of the Model 7001, select the
slot (CARD 1 or CARD 2) that you wish to install
the card in.
3. Referring to Figure 2-1 or Figure 2-2, feed the card
assembly into the desired slot such that the edges
of the relay card ride in the rails.
4. With the ejector arms in the unlocked position,
push the card assembly into the mainframe until
the arms engage into the ejector cups. Then push
both arms inward to lock the card all the way into
mainframe.
5. For the multi-pin connector card, install the screw
shown in Figure 2-1.
Notes:
1. Once a Model 701X series card is installed in the
mainframe, the Model 7001 automatically identi-
Þes it on power up. This ID allows the Model 7001
to conÞgure itself for proper operation.
2. When a Model 701X series card is installed, the slot
assignment (Card TYPE) is performed automati-
cally on power-up. A slot assignment change could
make a channel that is currently included in the
Scan List and/or a Channel Pattern unavailable.
The unavailable channel causes the Scan List and/
or the affected Channel Pattern to clear (see para-
graph 4.5.2 and 4.5.5 for details).
Card removal
To remove the card assembly, Þrst unlock it by pulling
the ejector arms outward, then pull the card assembly
out of the mainframe. Remember to handle the card as-
sembly by the edges and shields to avoid contamina-
tion that could degrade performance.
2-2
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
2.3

Non-701X series cards

WARNING
Turn off power from all instrumenta-
tion (including the Model 7001 main-
frame) and disconnect their line
cords. Make sure all power is re-
moved and stored energy in external
circuitry is discharged.
NOTE
If using a card that requires connec-
tions to be made on the card, make
sure your external circuitry is wired to
the card (as explained in the instruc-
tion manual for the card) before in-
stalling the card in the Model 7001
mainframe.
CAUTION
To prevent contamination to the
switch card that could degrade per-
formance, only handle the card by
the edges. If contamination occurs,
clean the card as explained in its in-
struction manual.
Card installation
Perform the following steps to install the card in the
Model 7001 mainframe:
1. Facing the rear panel of the Model 7001, select the
slot (CARD 1 or CARD 2) that you wish to install
the card in.
2. Using Figure 2-1 or Figure 2-2 as a guideline, feed
the card into the desired slot such that the edges of
the card ride in the rails.
3. With the ejector arms in the unlocked position,
push the card assembly into the mainframe until
the arms engage into the ejector cups. Then push
both arms inward to lock the card into the main-
frame.

Advertisement

Table of Contents
loading

Table of Contents