Mounting Screws For The Load Cells; Routing The Load Cell Cable - ABB PFEA 111 User Manual

Web tension systems with tension electronics
Table of Contents

Advertisement

Tension Electronics PFEA111/112, User Manual
Appendix D PFRL 101 - Designing the Load Cell Installation
D.8.2 Mounting Screws for the Load Cells
Measuring
direction
Transverse
direction
D.8.3 Routing the Load Cell Cable
D-12
The load cell shall be mounted with screws according to
The screws shall be tightened according to the manufacturer´s recommendations.
Screws with strength class 8.8 are sufficient for normal applications without large transverse
forces or overloads.
Screws with strength class 12.9 and a higher tightening torque are recommended in applications
where large transverse forces or overloads can occur.
Prior to mounting, check that the mounting surfaces are clean and flat, e.g. free of burrs and
other damage.
Table D-1. Mounting Screws
Load cell PFRL 101
A
B
C
D
The cable must be supported with clamps and routed to prevent force shunting through the
cable.
Table
NOTE
Screw dimension
M8 (5/16 UNC)
M8 (5/16 UNC)
M10 (3/8 UNC)
M12 (1/2 UNC)
D-1.
3BSE029380R0101 Rev C