Input Mains
The XMP 2600 operates at the following mains nominal voltages:
Voltage
Rating
Do not exceed this voltage range (nominal), as deterioration of performance or
damage to the unit is likely to occur. The following table shows the mains voltage
specification of the Power Supply:
Mains supply voltage fluctuations may not exceed +/-10% of the nominal voltage.
Power plug shall be fitted according to each national electrical code, and rated for a
current of at least 20A.
Do Not Open
The user, except when replacing modules, should not open the unit.
The Unit
Air Inlets And
The Power Supply is a forced air-cooled unit. Cooling is essential for the unit's
proper operation.
Air Outlet
The unit includes four air inlet openings located in the front panel, in both the left
and right sides near the front panel and in the top cover near the front panel. The air
outlet is in the rear panel (all over its surface).
Make sure there are no obstructions for the airflow, at least two centimeters (0.8")
away from these openings.
Obstructing these ventilation openings may cause fire and irreversible damage to the
unit.
Output
The Power Supply's outputs (at the unit's rear) may carry high voltages. Make sure
to handle them properly to avoid shock hazard.
Voltages
Unused outputs must be closed with load connectors (these connectors must include
sense lines connections).
Nonoperative
Do not operate the unit if it contains damaged or defective modules. Hazard or
damage may occur.
Modules
Damaged or defective units should be repaired by authorized personnel only.
f
170Vac - 265Vac nominal (45 to 66Hz)
120Vac nominal, up to 1KW output power (45 to 66Hz)
Nominal Mains
120Vac, Po<1KW
The XMP 2600 must be operated at the proper mains voltage. Before operating
the unit make sure the mains voltage complies with the range described above.
Noncompliance with the input voltage may cause hazard and damage to the unit!
Make sure the power cord is disconnected from the mains before opening the
unit. High voltages inside the unit may cause electric shock.
XMP 2600 Programming Manual
Nominal Current
230Vac
Iin<15A
Iin<15A
WARNING
WARNING
rev. 1.1