Activation - Xantrex XMP 2600 Programming Manual

Xantrex multiple output power system
Table of Contents

Advertisement

Recommendations on using the XMP 2600
Protection setup
Use the PROT command to setup the way Over Voltage and Over Current protection
values are assigned (automatically or manually).
If you have decided to use manual protection settings, use the OVSET and OCSET
commands to setup the desired protection values.
Select the Foldback scheme you wish to use (FOLD command) and the amount of
reprogramming delay you wish to apply (DLY command).
If you wish to use the window feature of the XMP 2600 (in order to receive warnings or
employ an Under Voltage or Under Current protection mechanism) enable the proper
window level thresholds using the WHIGH and WLOW commands.
Use the POLEN command to enable the XMP 2600 to process negative voltage setting
values.
Output settings
Setup the desired output values (VSET and ISET commands) and, optionally, the window
thresholds (VLOW, VHIGH, ILOW and IHIGH).
Remember that high window values can only be programmed to values above the
appropriate set values. Thus, for example, if you wish to use a window high current
threshold value lower than the limiting output value (ISET) you should program ISET to
the expected value, setup the ILOW and IHIGH values and then raise ISET to the limit
value.
Set the desired initial state of the output disconnect (DISC command) and polarity (POL
command) relays.
If the hardware SYNC signal is going to be used, use the SYNC command to tell the
Power Module when to generate a SYNC signal.
Setup the Power Module's reaction to a hardware or software trigger, using the TRIG
command.
Status reporting
If so desired, use the CMASK command to setup each Power Module's event filtering
mechanism. This mechanism enables you to limit the type of events reported by the
Power Modules to the XMP's main controller.

Activation

Once everything is setup properly, you are ready to activate the outputs of the XMP 2600.
Output on
For a Power Module's output to be active, the following conditions must be met:
To control the output of an individual Power Module, use the OUT command with the
module's number as a parameter.
10
The Main Converter is operating.
The XMP 2600 is globally "turned on".
The Power Module is turned on.
The Power Module does not report any fault.
The output disconnect relay is closed.
XMP 2600 Programming Manual
Usage Guidelines
rev. 1.1