Xantrex XMP 2600 Programming Manual page 62

Xantrex multiple output power system
Table of Contents

Advertisement

Commands and Queries Reference
GLBL
GLBL?
GPIB
GPIB?
GRP
inform the XMP 2600 Main Controller on shut-down behavior of the
installed Power Modules.
Syntax
Parameters
Category
return the current Global Shut-down set-up of all Power Modules. The
response is made up of two decimal numbers (each in the range 0 to
255) being the high (sent first) and low parts of a 16 bits word. Each
bit in the response represents the Global set-up of a Power Module
(see GLBL command for details on bit values).
Syntax
Parameters
Category
set-up the operating mode of the XMP 2600. Re-calculate <rqs>.
Syntax
Parameters
Category
return the current operating mode of the XMP 2600. The value
returned is a decimal number in the range 0 to 1.
Syntax
Parameters
Category
define the members of the group of Power Modules to be shutdown
when one (or more) is faulty.
Syntax
Parameters
Category
XMP 2600 Programming Manual
GLBL <int>,<int>
the two integers are (respectively) the high and low
parts of a 16 bits word where each bit represent a
Power Module. If the bit value is "0" then that
Power Module will not globally shut-down when
faulty. If the bit value is "1" then this Power Module
will cause a global shut-down when it is faulty.
Works in combination with the GRP command.
System
GLBL?
none
System
GPIB <int>
the mode the XMP 2600 is to be operating in: a
value of "1" is for the IEEE 488.2 compatible mode.
System
GPIB?
none
System
GRP <int>,<int>
the two integers are (respectively) the high and low
parts of a 16-bit word where each bit represents a
Power Module. If the bit value is "0" then that
Power Module is not part of the group. If the bit
value is "1" then this Power Module will be
shutdown when any of the group's members is
faulty.
System
rev. 1.1
Alphabetical reference
Type: Sequential
Type: Sequential
Type: Sequential
Type: Sequential
Type: Sequential
49