Calibrate O 2 Sensor - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Calibrate O
Sensor
2
To calibrate the oxygen sensor correctly, make sure
it is exposed only to room air during the entire cali-
bration period.
To avoid leakage, remove the oxygen sensor
assembly from the inspiratory valve dome, and
insert the valve dome plug into the dome.
1 Press the »Calibrate O
the Setup window.
The calibration instruction window replaces the
Setup window. Follow the directions provided:
– Remove O
sensor and expose to room air for
2
2 minutes
– To start O
Calibration press rotary knob
2
– Observe Calibration status in O
– Reinsert O
Sensor after successful calibration.
2
Instructions for Use Fabius plus SW 3.n
Sensor« soft key in
2
data window
2
Configuration
1
145