Ending Operation
1 Press and confirm the »
Monitoring and alarms are turned off and the venti-
lator stops.
2 Turn vaporizers off by turning the handwheel to
»0« until the button engages.
Turn off fresh-gas flow. Sleep mode will activate
2.5 minutes after fresh gas is turned off.
Close the cylinder valves.
WARNING
Damage of materials and health risks
Never switch off fresh-gas flow before the
vaporizer is switched off. A vaporizer must
never be left switched on without a fresh-gas
flow, because high-concentration anesthetic
vapor may leak into machine lines and ambi-
ent air.
When Fabius plus is not in use
If Fabius plus is not used for an extended period:
Unplug the medical gas hoses from the wall
supply points of the central gas supply.
Close the cylinder valves on the reserve gas
cylinders.
NOTE
Leave the Fabius plus plugged into mains power in
order to charge the battery.
Instructions for Use Fabius plus SW 3.n
« key (Standby).
Operation
1
2
97