For Your Safety and that of Your Patients
WARNING
Do not apply unregulated suction to the pati-
ent circuit when using the device.
WARNING
To avoid electrical shock hazard: Due to the
risk of electrical shock, do not remove any
component cover. Refer any servicing to
DrägerService. Use only hospital-grade
grounded electric outlets and power cord.
This device is to be used only in rooms with
line power installations complying with natio-
nal safety standards for hospital patient
rooms (e.g. IEC 60601-1 "Safety of Medical
Electrical Equipment"). Make sure the external
equipment is hospital-grade (regarding natio-
nal regulations) grounded before connecting
the equipment. Disconnect the power supply
from the electrical outlet before cleaning or
servicing. Let it dry completely before recon-
necting it to the electrical outlet. Always
ensure that the clamp for the power cord at the
power supply end is tight, thus preventing an
accidental disconnect from the unit. Do not
connect additional external equipment other
than equipment specified by Dräger.
CAUTION
Communications with external equipment may be
temporarily affected by electromagnetic interfe-
rence due to the use of electrosurgical equipment.
CAUTION
Do not use Fabius plus during magnetic reso-
nance (MRI, NMR, NMI).
Device operation may be affected, thus placing
the patient at risk.
12
CAUTION
Risk of physical injury
To avoid physical injury, pay special attention to
edges, moving parts and corners when working
with:
– drawers,
– the ventilator module,
– the writing tray,
– swivel arms for mounted devices,
– gas cylinders,
– vaporizer units,
– CLIC absorbers and CLIC adapters,
as well as other accessories.
CAUTION
Risk of pinching fingers or breathing hoses and
objects falling down
If the writing table is not engaged correctly,
objects can fall down and fingers or breathing
hoses can be pinched.
CAUTION
Risk of device failure
If the anesthesia workstation is operated when til-
ted, components may be damaged or may func-
tion improperly.
Do not operate the anesthesia workstation if it is
tilted more than 1°.
NOTE
Software must be installed by qualified personnel.
We recommend to contact DrägerService for soft-
ware installation.
NOTE
If the error-free state of the protective earthing con-
ductor or its connection to the device is doubtful,
the device must be operated using the internal
power supply (battery).
Instructions for Use Fabius plus SW 3.n