Connecting the Reserve Gas Cylinders
for N
O, AIR and O
2
Threaded Connectors)*
CAUTION
Risk of explosion
Do not oil or grease the O
pressure regulator.
CAUTION
Even if the gas supply is connected to a medical
gas pipeline, the cylinders should remain on the
device in reserve.
CAUTION
Risk of patient injury
Do not connect the cylinders to the pressure inlet
without pressure reducers listed in the accesso-
ries list 9052267.
NOTE
If cylinder valves are leaky or difficult to open or
close, they must be repaired in accordance with
the manufacturer's specifications.
NOTE
The cylinder valves must only be opened or closed
manually. Never use tools.
NOTE
Reserve gas cylinders should remain closed when
not in use. Risk of unintentional emptying of cylin-
ders.
*
optional
Instructions for Use Fabius plus SW 3.n
(for Cylinders with
2
cylinder valves and O
2
2
Assembly
53