O 2 Monitoring Disabled; O 2 Low Supply Alarm - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Monitoring
O
Monitoring Disabled
2
The following oxygen monitoring functions are dis-
abled if the Fabius plus is configured by
DrägerService to run using the O
abled option.
– "Oxygen Monitoring Window" on page 112
– "Setting Oxygen Monitoring Alarm Limits" on
page 112
– "Calibrating the Oxygen Sensor" on page 113
– Fabius plus-generated inspiratory O
sensor alarms.
NOTE
If internal FiO
monitoring is deactivated, external
2
FiO
monitoring must be available (according to
2
ISO 21647).
NOTE
The message »No Integrated O
displayed in the oxygen monitor window when O
monitoring is disabled.
O
Low Supply Alarm
2
If the O
pipeline supply pressure drops below the
2
minimum pressure permitted (approximately 20 psi
or 1.4 kPa x 100) an additional »O2 SUPPLY
LOW« alarm is generated and the LED (1) starts
blinking.
If the alarm occurs in Standby mode and the user
changes to a ventilation mode, an continuous alarm
tone sounds for seven seconds.
116
Monitoring Dis-
2
and O
2
2
Monitoring« is
2
2
Instructions for Use Fabius plus SW 3.n
1