Serial Interface

Type:
Connector:
Baud Rates:
Parity:
Data Bits:
Start Bits:
Stop Bits:
Protocol:
Instructions for Use Fabius plus SW 3.n
RS-232-C
9-pin Sub D Female, Shield on housing
Pin 2 TXD
Pin 3 RXD
Pin 5 GND
Galvanic Isolation 500 V
1200, 2400, 4800, 9600, 19200, 38400
Odd, Even, None
7 or 8
1
1 or 2
Vitalink, MEDIBUS