Electrical Safety Conformance; General Safety Standards For Anesthesia; Freedom From Latex; Ventilator - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Technical Data

Electrical Safety Conformance

Conform to:
IEC 60601-1
General Safety Standards for
Anesthesia
IEC 60601-2-13
ISO 8835-2

Freedom from latex

Fabius plus is latex-free!
Latex-free breathing bags and breathing hoses
must be used for latex-free use.

Ventilator

The anesthetic ventilator complies with the require-
ments of ISO 8835-5.
Control Inputs Ranges
P
MAX
V
T
V
(SIMV/PS)
T
f
T
:T
I
E
T
:T
IP
I
PEEP
P
INSP
200
Pressure limiting
Tidal volume
Tidal volume
Breathing frequency
Inspiration/expiration ratio
Inspiration pause
End-expiratory pressure
Inspiratory pressure
15 to 70 cmH
O (1 cmH
2
(15 to 70 hPa (1 hPa resolution))
(setting must be at least 10 cmH
PEEP; and in SIMV/PS mode, the P
Δ
also be greater than
20 to 1400 mL (10 mL resolution)
20 to 1100 mL (10 mL resolution)
4 to 60 bpm (1 bpm resolution)
(4 to 60 1/min (1/min resolution))
4 : 1 to 1 : 4
0 % to 50 % (1 % resolution)
0 to 20 cmH
O (1 cmH
2
(0 to 20 hPa (1 hPa resolution))
5 to 65 cmH
O (1 cmH
2
(5 to 65 hPa (1 hPa resolution))
(setting must be at least 5 cmH
PEEP)
Instructions for Use Fabius plus SW 3.n
O resolution)
2
O (10 hPa) above
2
setting must
MAX
P
+PEEP)
PS
O resolution)
2
O resolution)
2
O (5 hPa) above
2