Calibrate O 2 Sensor - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Calibrate O
Sensor
2
Press the »Calibrate O
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
Follow the instructions on the screen:
– Remove O
sensor and expose to room air for
2
2 minutes
– To start O
Calibration press the rotary knob
2
When the calibration begins, the instructions
are removed, and a »O
progress« message is displayed above the
Standby soft keys.
When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »O
Sensor Calibration completed –
2
reinsert O
sensor« or »O
2
tion Failed«.
Troubleshooting O
Sensor Calibration Failure
2
If the O
sensor cannot be calibrated, replace the
2
O
capsule in the O
sensor housing (see
2
2
page 185).
If the O
sensor still cannot be calibrated, call
2
DrägerService or your local authorized service
organization.