Removing the Compact Breathing

System

Remove all breathing hoses.
Disconnect the breathing bag extension and
bag by loosening the two thumb screws.
Remove the ventilation hose.
Remove the fresh-gas hose from the breathing
system.
Remove the scavenger hose.
Remove the flow sensor cable.
Remove the O
sensor cable.
2
Remove the breathing pressure cable.
Detach the APL-bypass and the PEEP/P
lines from the breathing system and from the
side of the machine.
Remove the absorber (for complete instruc-
tions, see page 185).
Remove the compact breathing system.
1 Remove plug for the inspiration dome.

Removing the Inspiratory Valve

2 Unscrew the retaining nut.
3 Remove the inspection cap.
4 Extract the valve disc.
5 Remove the gasket.

Removing the Expiratory Valve

6 Unscrew the retaining nut.
7 Remove the inspection cap.
8 Extract the valve disc.
9 Remove the gasket.