Using the external fresh-gas outlet as a
common gas outlet*
WARNING
Risk of insufficient gas supply to the patient
Non-rebreathing systems are suitable and
intended only for manual ventilation or spon-
taneous breathing and may only be connected
to the external fresh-gas outlet.
WARNING
Risk of misinterpreting the measurements
The measurements and alarms are displayed
on the breathing monitor. They do not corres-
pond to the measurements for the patient con-
nected to the external fresh-gas outlet.
– The measurements (O
and alarms displayed on the breathing
monitor correspond to the measurements
on the Compact Breathing System (COSY).
Switch to the Standby mode when using the
external fresh-gas outlet.
WARNING
Danger to the patient
There is no monitoring of pressure, volume or
O
for the external fresh-gas outlet.
2
An external monitoring unit must be used to
perform the required monitoring of O
anesthetic gas, volume and pressure.
*
optional
98
, pressure, volume)
2
, CO
,
2
2
Instructions for Use Fabius plus SW 3.n