Reinstalling The Suction System; Reconnecting The Suction System - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Reinstalling the Suction System

Reconnecting the Suction System

The optional aspiration system for the Fabius plus
consists of an suction regulator and a suction bot-
tle. The suction regulator is attached to a holder,
which is fastened on the side channel of the anes-
thesia device. The suction bottle desired by the
customer is attached to a separate swivel rail on the
side channel.
Attach the carrier arm of the suction system on
the side channel on the side of the anesthesia
device.
1 Mount the suction regulator onto the bracket.
Reprocess the suction bottle according to the
instructions for use included with the bottle.
2 Attach the suction bottle on the swivel rail.
Depending on the aspiration system used:
With use of AIR or O
as driving gas:
2
Connect the air connection hose of the suction
system on the air exhaust of the gas supply
block (optional) or directly on the gas supply
line.
For vacuum aspiration:
Connect the vacuum hose of the suction system
directly on the gas supply line.
Make sure that the suction system is ready for oper-
ation according to the included Instructions for
Use.
WARNING
The aspiration system must only be used in
the »Man/Spont« mode or if the Y-piece is not
connected.
Instructions for Use Fabius plus SW 3.n
1
2