Auxiliary Fresh-Gas Outlet*
Auxiliary fresh-gas outlet is an option for external
non-rebreathing systems, e.g. Magill, Kuhn,
Waters, Bain (for use with anesthesia machines).
WARNING
Danger to the patient caused by too high pres-
sures
To ensure the correct inspiratory pressure at
the inspiratory port, use only non-rebreathing
systems with a reservoir bag conforming to
ISO 5362 and/or a pressure relief valve.
WARNING
Risk of insufficient gas supply to the patient
Non-rebreathing systems are suitable and
intended only for manual ventilation or spon-
taneous breathing and may only be connected
to the external fresh-gas outlet.
WARNING
Risk of misinterpreting the measurements
The measurements and alarms are displayed
on the breathing monitor. They do not corres-
pond to the measurements for the patient con-
nected to the external fresh-gas outlet.
– The measurements (O
and alarms displayed on the breathing
monitor correspond to the measurements
on the Compact Breathing System (COSY).
Switch to the Standby mode when using the
external fresh-gas outlet.
WARNING
Danger to the patient
There is no monitoring of pressure, volume or
O
for the external fresh-gas outlet.
2
An external monitoring unit must be used to
perform the required monitoring of O
anesthetic gas, volume and pressure.
*
optional
Instructions for Use Fabius plus SW 3.n
, pressure, volume)
2
, CO
,
2
2
NOTE
The connected system (e.g. Magill, Kuhn, Waters,
Bain) must be operated in accordance with the
respective associated Instructions for Use.
System Overview
27