Reinstalling The Scavenger System; Reconnecting The Anesthetic Gas Receiving System (Ags) - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Reinstalling the Scavenger System

Reconnecting the Anesthetic Gas Receiving
System (AGS)
Every Anesthetic Gas Scavenging System (AGSS)
used on the Fabius plus must follow ISO standard
8835-3.
The scavenging system is used with vacuum
waste-gas disposal systems.
The AGS is not applied as an independent system.
It is used as one of the three components of the
AGSS.
WARNING
Risk of patient injury.
If the side openings of the receiving system
are blocked, negative pressure may result in
the breathing system and the patient's lungs.
Always make sure the side openings of the
receiving system are not blocked.
NOTE
Remove the socket from the scavenger hose
before connecting.
NOTE
The scavenger hoses must not be pinched,
kinked, or blocked in any manner.
Hook receiving system with the slots over the
appropriate pins of the basic device and allow it
to slide down into place.
1 Connect the scavenging hose to the relevant
socket of the receiving system.
Connect the connector of the scavenging hose
to the terminal unit of the disposal system.
2 Close the connection that is not used with a
screw cap.
Push the transfer hose on the designated
socket.
Connect the other end of the transfer hose to
the waste-gas port located on the underside of
the breathing system.
Instructions for Use Fabius plus SW 3.n
1
2