Procedure
1. Select [Save (.fcs)].
2. Choose a Storage folder and insert Sample name.
3. Click [Save].
For more information, please refer to chapter 13.3 Data management.
7.5 Performing intermediate cleaning
A cleaning mode is available to clean the fluidic system of the device.
Intermediate Cleaning: Choose the clean symbol in the right lower corner (Device
Controls).
INFORMATION
Intermediate Cleaning
Intermediate Cleaning reduces the possibility of cross contamination and
background signals.
Requirements
•
Sample tube with 2 ml Cleaning Solution
•
Sample tube with 2 ml Sheath Fluid
Procedure
1. Connect a Sample tube filled with Cleaning Solution.
2. Click [Intermediate Cleaning].
3. Connect a Sample tube filled with Sheath Fluid.
4. Click [Intermediate Cleaning].
7.6 Shutdown procedure
7.6.1 Shutdown cleaning
A shutdown cleaning procedure is available to clean the fluidic system of the device.
Requirements
•
Sample tube with 2 ml Cleaning Solution
•
Sample tube with 2 ml Decontamination Solution
•
Sample tube with 2 ml Sheath Fluid