6 Operation
INFORMATION
Intermediate Cleaning
Intermediate Cleaning reduces the possibility of cross contamination and
background signals.
Requirements
Sample tube with 2 ml Cleaning Solution
Sample tube with 2 ml Sheath Fluid
Procedure
1. Connect a Sample tube filled with Cleaning Solution.
2. Click [Intermediate Cleaning].
3. Connect a Sample tube filled with Sheath Fluid.
4. Click [Intermediate Cleaning].

6.5 Shutdown procedure

6.5.1 Shutdown cleaning

A shutdown cleaning procedure is available to clean the device. Shutdown procedure:
Choose the clean symbol in the upper part below the menu bar (File Operations).
Requirements
Sample tube with 2 ml Cleaning Solution
Sample tube with 2 ml Decontamination Solution
Sample tube with 2 ml Sheath Fluid
Procedure
1. Select the [Shutdown Cleaning mode].
2. Press [Start].
3. Follow the instructions viewed.
4. The software will shut down automatically once the procedure is finished.
30
Figure 10: Intermediate Cleaning
Figure 11: Selecting Shutdown Cleaning Mode
CyFlow™ Cube 6 V2m | Operating Manual | February 2020