Overview; Cyview™ Script; Fcs Express™ Template; Quality Check Material - SYSMEX CyStain VitalCount Manual

Table of Contents

Advertisement

3 Overview

3.1 CyView™ script
The Quality Check procedure is integrated into a CyView™ configuration file (*.cvc85) -
called a CyView™ script. To access the QC procedure, the "QC"-Mode within the respective
CyView™ script can be activated.
3.2 FCS Express™ template
FCS Express™ templates are used to automatically analyse the results of QC
measurements. The templates can also be used to analyse data from QC measurements
manually. The results can be exported.

3.3 Quality Check material

The QC procedure uses the following material(s). For further information please refer to the
device manual.
NOTICE
Count Check Beads green specificity
This QC procedure was designed for the use with Count Check Beads
green (REF 05-4011_R). Similar products such as Count Check Beads
green (REF 05-4011) are not suitable for use.
Always use the recommended reagents.
Light source
Item
Blue Laser
Count Check Beads green 50 tests
CyStain™ VitalCount QC Procedure | QC Manual | June 2021
Content
3 Overview
REF
05-4011_R
5