Your 220 indicator has been thoroughly tested and
calibrated before being shipped to you. If you
received the indicator attached to a scale,
calibration is not necessary. If the indicator is
being connected to a scale for the first time or
recalibration is necessary for other reasons,
proceed as indicated.
The calibration switch is located on the enclosure
rear panel. You may gain access to this switch
simply by removing the calibration switch access
screw on the rear panel. Refer to Figure No. 8.
During the setup and calibration process it is necessary to enter operational parameters via the
220's keypad. Pressing the ENTER key without entering a new value will retain the current
setting and advance the 220 to the next prompt. To change a setting, enter a new value and
press the ENTER key. This will save the new value and advance the 220 to the next prompt.
Pressing the ASTERISK key will "backup" to the previous prompt or return to normal operation.
CAUTION: The membrane keypad is not to be operated with pointed objects
(pencils, pens, fingernails, etc). Damage to keypad resulting from this practice will
NOT be covered under warranty.
Enter Setup Mode
To enter the setup mode, with the indicator ON, insert a small screwdriver or other tool through the
calibration switch access hole on the rear panel. Press and release the calibration switch. The
menu SETUP will be displayed. Continue to press and release (or hold) the switch to rotate through
the beginning point for entering the setup mode.
SETUP
ANALOG
CALIBRATE
SERIAL
PRINT TABS
FINE SPAN
HIGH RES
KEY LOCK OUT
ANALOG OUT
If you press the ENTER key at the SETUP prompt, you may proceed through to the next section (up
to and including FINE SPAN) by pressing the ENTER key.
NOTE! Setup may be interrupted at any time. ALL data previously entered and
finalized with the ENTER key will be retained in the non-volatile memory.
Pressing the calibration switch at any prompt will return you to the SETUP menu. To exit setup,
press the ASTERISK key with any of the above menu selections displayed or cycle power at any
time (press the ON/OFF key twice).

SETUP AND CALIBRATION

Setup Mode (starts at USA= prompt)
Analog to Digital Filtering (starts at FILTER= prompt)
Calibration (starts at CAL WT 1= prompt)
Serial Input/Output (starts at BAUD 1= prompt)
Print Tab Settings (starts at PORT= prompt)
Fine Span Adjustment
Display High Resolution Weight Mode
Key Lock Out Function
Analog Output Option (If option board is installed, Calibration
of 10 volt or 4 to 20 mA Analog Output)
15
Calibration Switch
Access Screw
Figure No. 8