Your 185 or 185B Weight Indicator has been thoroughly tested and calibrated before being
shipped to you. If you received the indicator attached to a scale, calibration is not necessary. If
the indicator is being connected to a scale for the first time or recalibration is necessary for
other reasons, proceed as indicated.
Calibration of the indicator is accomplished entirely by the
keypad. To enter setup and calibration, the calibration
jumper must be removed and re-installed while the
indicator is on. The calibration jumper is located on the
main printed circuit board. Refer to Figure No. 11 for the
location of the jumper. Loosen the four (4) Phillips screws
securing the rear housing to the front housing, and remove
the rear housing to access the calibration jumper.
During the setup and calibration process, it will be
necessary to enter operational parameters via the
indicator's keypad.
 When a prompt for a value is displayed, press the TARE  key to display the current setting.
 Pressing the TARE  key will cause the data entered or displayed to be retained and the
indicator to advance to the next prompt.
 The functions of numeric keys are replaced by using the UNITS/ and the Fn/▲ keys.
 The cursor location is identified by the blinking character and can be advanced to the left to
the next position by pressing the UNITS/ key.
 Pressing the Fn/▲ key will change the blinking character to the next value. Continue to press
this key to "toggle" between the different available values for the setup parameter.
 Pressing the Fn/▲ key when a setup parameter (not a parameter value) is displayed, will
"backup" to the previous prompt.
The membrane keypad is not to be operated with pointed objects (pencils,
pens, fingernails, etc.). Damage to keypad resulting from this practice is
NOT covered under warranty.
Begin Setup and Calibration:
1. With the rear panel removed and the indicator ON, remove the calibration jumper J3.
2. The SETUP prompt will be displayed.
3. Re-install the calibration jumper and press the TARE  key.
4. The rdisp= prompt will be displayed.
5. The indicator is now ready for setup and calibration.
rdisp= (Remote Display)
This setting will turn the 185 or 185B indicator into a remote display using the serial port. If
the 185 or 185B is connected to a Cardinal 200 series indicator, it becomes a functional
remote keypad and display.
Press the TARE  key to view the current setting. To keep the current setting, press the
TARE  key again. Otherwise, press the Fn/▲ key to scroll through the available values.
When the desired value is displayed, press the TARE  key to select the displayed value
and proceed to the next prompt.
Available values are:
YES
Remote Display Function Enabled
If rdisp= YES is selected, an additional prompt, tYPE= will be displayed.
8400-0181-OM Rev C  185 and 185B Installation & Technical

SETUP AND CALIBRATION

no
Figure No. 11
Remote Display Function Disabled
15