CONTINUOUS DATA OUTPUT FORMATS, CONT.
GEDGE 1650 FORMAT 4
Beginning with version 2.6.1 of the firmware, the GEDGE 1650 Format 4 has been added to
the Continuous Data Output Formats. In setup, select type 10 for the GEDGE format.
At the TdAte prompt, enter the format for the date and time (0-4).
At the StrtC prompt, enter the decimal value for the start character (0-126)
At the EbLoC prompt, enter the decimal value for the end block character or data
separator (0-126).
At the dbUb prompt, enter the decimal value for the device number (0-99).
Format
<Bg><DT><BD><No><BD><W1><BD><D2><D3><D4><D5><D6><D7><BD><En>
Where,
Bg
is the starting character, if 0 then it is not transmitted.
DT
is the Date and Time the following BD character is output only if the
date/time is output.
BD
Block division character.
D2
Displayed weight identity. G=gross, N=net
D3
M scale is in motion, S scale is stable.
D4
Inscale (0>=weight<=capacity) Overcap U=below zero
D5
Z=Center-of-Zero, otherwise ASCII space
D6
E=Tare not equal to 0, otherwise ASCII space
D7
Always an ASCII space
En
End of transmission character
No
Device Number 01 thru 99. If set to 0 above, then nothing is output.
W1
Displayed weight.
4
200, 205, 210 and 215 Rev 2.6.1 Addendum