Your 758C Indicator has been thoroughly tested and calibrated before being shipped to you. If
you received the indicator with a scale, calibration is not necessary. If the indicator is being
connected to a scale for the first time or recalibration is necessary for other reasons, proceed
as indicated.
Calibration and Setup of the indicator is accomplished entirely by
the keypad. To enter the setup and calibration mode:
1. With the power off, remove the 2 screws from the right
end cap (as viewed from the indicator back) and remove
the end cap.
2. Locate the calibration jumper J2 on the circuit board and
remove the jumper. See Figure No. 9.
3. With the jumper removed, press and hold the
LOCK/RELEASE key and then press the ON/OFF key.
4. The indicator will perform a display test (turn on all digits),
show the software version for a few seconds and then
change to show SETUP.
5. Release the LOCK/RELEASE key.
6. The display will change to setgC.
7. The indicator is now ready for setup and calibration
During the setup and calibration process it is necessary to enter operational parameters via the
indicator's keypad. The following keys and the numeric keypad are used to input the values.
ON/OFF
PRINT/ENTER
1. Pressing the PRINT/ENTER key without entering a new value will retain the current
setting and advance to the next prompt.
2. Some prompts require only a YES or NO selection.
3. Other prompts require a numeric value to be entered. Use the numeric keypad to enter
the new value.
4. In either case, pressing the PRINT/ENTER key after changing a setting will save the new
value and advance to the next prompt.
CAUTION! The membrane keypad is not to be operated with pointed
objects (pencils, pens, fingernails, etc). Damage to keypad resulting from
this practice will NOT be covered under warranty.
setgC (SET GRAVITY CONSTANT)
This indicator is equipped with an acceleration of gravity function which means that it can be
calibrated in one location and then adjusted to match the acceleration of gravity at the
location where it will used.
With display showing setgC, press PRINT/ENTER key to show current setting. If setting
displayed is acceptable, press PRINT/ENTER key to save it. Otherwise use numeric keys
0/NO or 1/YES and enter a new setting. Press PRINT/ENTER key to save it.
0 = Gravity Constant is NOT used
NOTE: If you select 1 (Set Gravity Constant) two additional prompts will be displayed.

SETUP AND CALIBRATION

NO
1 = Set Gravity Constant
8555-M210-O1 Rev R  758C Owner's
Calibration
Jumper J2
Figure No. 9
YES
19