SETUP AND CALIBRATION, CONT.
A - d (A - d?) - Analog to Digital Filtering
dFLt= (Digital Filtering)
With A - d (A - d?) displayed, press the ENTER key. The display will change to dFLt=.
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key to save it. Otherwise, using the numeric keys enter the new setting, then
press the ENTER key to save it. Allowable values are: 0, 1, 2 or 3. Note, that if you select 3
(Custom Filtering) two additional prompts will be displayed.
dFLt=
0
Disabled – NO Filtering
1
MINIMAL FILTERING (sample rate = 2)
2
MODERATE FILTERING (sample rate = 1)
3
CUSTOM FILTERING
NOTE! The prompts, F= (Filter Level) and b= (Break Range) will only be displayed if you
selected 3 (Custom Filtering) for the dFLt= (Digital Filtering) prompt.
F= (Filter Level)
Press the ENTER key to show the current setting for the filter level. The filter level is a number
from 1 to 99 that corresponds to the level of filtering with 99 being the greatest filtering and 1
the least. To accept the value displayed, press the ENTER key, otherwise, use the numeric
keys to enter a new value then press the ENTER key to save it.
b= (Break Range)
Press the ENTER key to show the current setting for the break range. The break range is a
number from 1 to 255 that corresponds to the number of division change to break out of the
filtering. Press the ENTER key to keep the displayed value or use the numeric keys to enter a
new value and press the ENTER key to save the new setting. Note that entry of a 0 disables
this feature.
Sr= (Sample Rate)
Press the ENTER key to show the current setting for the sample rate. The value displayed is
the sample rate in samples per second. Press the ENTER key to save the displayed value or
use the numeric keys to enter a new value (1 to 100) and press the ENTER key to save it.
UnS= (Motion Range)
Press the ENTER key to view the current setting for the range of motion detection. If the
displayed value is acceptable, press the ENTER key to save it. Otherwise, use the numeric
keys to enter the new range (the number of divisions of change permitted before indicating
unstable), then press the ENTER key to save the new setting. Allowable range values are: 0
through 99 divisions.
SC= (Stable Count)
Press the ENTER key to view the current setting for the number of consecutive stable weight
readings before indicating stable weight. This helps filter weight readings for stability for use
with Auto Print on Balance, or and anything trying to capture stable weight. If the displayed
value is acceptable, press the ENTER key to save it. Otherwise, use the numeric keys to enter
a new value and press the ENTER key to save the new setting. Allowable values for the
stable count are: 3 through 255.
8200-M411-O1 Rev D  210/215 Installation & Technical