Dentsply Sirona Intego Ambidextrous Care And Maintenance

Hide thumbs Also See for Intego Ambidextrous:

Advertisement

Quick Links

Intego Ambidextrous
A continuously updated list of approved care, cleaning, and disinfecting agents can be downloaded from the Dentsply Sirona internet website under REF 59 70 905:
www.dentsplysirona.com/manuals
A
B
Purge
C
D
max.
Purge
E
134°C 274°F
3 min
A
TS dentist element
CS dentist element
Grease
For more details, refer to the operating instructions.
The national requirements and recommendations for hygiene and disinfection must be observed (e.g. RKI, ADA, CDC etc).
After each patient
Clean/disinfect surfaces
CAUTION!
Use only care, cleaning, and disin-
fecting agents approved by Dentsply Sirona!
• Disinfect surfaces and upholstery
• Disinfect user interfaces
• Care, disinfection/sterilization of treatment
instruments (see the operating instruc-
tions of the respective treatment instru-
ments)
• Clean and disinfect/sterilize the compo-
nents of the ApexLocator
• Disinfect instrument holders
• Disinfect handles
• Disinfecting the tray
• Disinfect connectors for suction hose
cleaning
• Maintain and clean the operating light (see
LEDlight, LEDlight Plus or LEDview Plus
operating instructions)
Clean/disinfect the water paths and
suction system
CAUTION!
Use only care, cleaning, and disin-
fecting agents approved by Dentsply Sirona!
A)
Clean the gold trap
CAUTION! Dispose of amalgam residues in
an environmentally correct manner!
B)
Clean/disinfect the (removable) cuspidor
bowl (special cleaning agent))
C)
Purge the water lines of used dentist
element instruments (Purge function)
D)
Clean the suction hoses
E)
Sterilize suction handpieces
To prevent reflux, please use the suction
cannula with supplementary air intakes.
Weekly
Clean and care for upholstery
CAUTION!
Use only cleaning and care
agents for upholstery approved by Dentsply
Sirona!
Follow the instructions for use of the
cleaning and care agents exactly.
Prevent agent FD360 from contacting
chrome-plated surfaces.
Cleaning the bottom surface of the
mount (A) for the backrest component
Clean the foot switch
Clean the underside of the foot switch with a
moist cloth (water).
If the "Chemical hose cleaning" option
is not available:
Clean drainage lines
The cleaning agents used for chemical
cleaning of the hoses and of the suction
system are identical (see REF 59 70 905).
TS dentist element
Change the cotton wool roll on the turbine
hose
CS dentist element
Change oil collector
Lubricating suction handpieces
Separate suction handpieces and grease the
O-rings after thermal disinfection or steriliza-
tion as well as once a week.
Treatment center care and cleaning plan (not valid for USA)
A
Auto
A
B
D
C
San
Disposal
Daily
If the "Chemical hose cleaning" option
is not available:
Disinfect the suction hoses using the
cleaning adapters (A)
• daily at noon and
• and at the end of the work day.
Automatically purge water lines (Auto-
Purge function)
Insert the couplings into their corresponding
adapters prior to the beginning of and at the
end of the work day:
• White - turbine
• Green - BL E motor
• Blue - BL ISO E motor (ISO interface)
• Red - SiroSonic L scaler
• Yellow - Sprayvit E, water on right button
• Orange - Sprayvit E, water on left button
Sterilize silicone mats and handle
covers
Thermally disinfect the tray and cup
holder
Disinfect dentist and assistant element
Empty the central suction sieve
Wear protective gloves when emptying the
central suction sieve!
Collect amalgam residues in a closed
container with water.
CAUTION!
Dispose of amalgam residues in
an environmentally correct manner.
Clean emptied suction sieve under running
water.
Monthly or as required
Refilling the cleaning agent tank
"chem" appears on the user interface: Refill
the cleaning agent as soon as possible for
chemical suction hose cleaning (optional).
Clean the filter insert of the standard wet
suction
On the Start screen, press the "Extract
residual water" button before removing the
filter for cleaning.
Change filter (C) for water (A) and air (B)
Empty the sediment tank (D) (optional)
Wear protective gloves when emptying the
sediment tank!
Microbiological water test
• At regular intervals according to results
and
• after periods of disuse > 1 week.
Sanitize the treatment center
For reduction of the biofilm within the fresh
water carrying systems of the treatment
center
• White - turbine
• Green - BL E motor
• Blue - BL ISO E motor (ISO interface)
• Red - SiroSonic L scaler
• Yellow - Sprayvit E, water on right button
• Orange - Sprayvit E, water on left button
Cleaning the suction hoses after sanitation.
Replace amalgam rotor
IMPORTANT!
The rotor must be replaced
after 12 months at the latest!
1. "Amalg" appears on the user interface.
Conduct disinfection.
2. Only with the unit switched on: remove
amalgam rotor.
3. Fill in collected amalgam residues, close
amalgam rotor.
4. Remove amalgam rotor from lower part
and dispose of in an environmentally
correct manner.

Advertisement

loading

Summary of Contents for Dentsply Sirona Intego Ambidextrous

  • Page 1 Intego Ambidextrous Treatment center care and cleaning plan (not valid for USA) A continuously updated list of approved care, cleaning, and disinfecting agents can be downloaded from the Dentsply Sirona internet website under REF 59 70 905: www.dentsplysirona.com/manuals After each patient...
  • Page 2 Intego Ambidextrous Conversion Preparation for optional equipment variants Converting the treatment center 1. Positioning the dentist element 2. Positioning the light 3. Starting the conversion process 4. Performing the conversion process Press pedal and move water unit forward. Acoustic signal sounds.