4.7.4

Stand-alone water supply

If the water quality of the public drinking water supply is not suitable for
treating patients, the treatment center can be operated with a self-
sufficient water supply.
In this case, food-grade water is mixed with the agent for disinfecting
the water paths in a 100:1 ratio (1 liter of water, 10 ml of the
disinfectant) and filled into the fresh water bottle or disinfectant tank of
the water unit. The disinfectant will reduce the microbial growth in the
water paths.
Please note that drinking water, e.g., water from bottles, must never be
filled into the disinfectant tank of the Comfort water unit (due to the
minerals they contain). Aqua purificata or aqua destillata can be used
as distilled water. Water from bottles can be used without any
restrictions for the fresh water bottle.
Regularly sanitize the water paths (every 4 weeks) and always after
switching to the public drinking water supply from the stand-alone water
supply (see "Sanitization"  [→ 293]).
The stand-alone water supply functionality of the Intego is only available
if the treatment center is equipped with a fresh water bottle. The Intego
Pro treatment center must be equipped with the integrated water
disinfection system.
In every mode, the cuspidor is flushed from the public drinking water
supply.
CAUTION
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one's
health.
➢ Dentsply Sirona recommends never operating the treatment
center without the agent for disinfecting the water paths.
➢ Mix fresh water for the self-sufficient water supply every day. At
the end of the work day, the fresh water bottle must be emptied
and remaining water must be flushed out of the disinfectant tank
using the tumbler filler.
➢ Check the amount of bacteria in the water of the treatment center
at regular intervals. In particular, if no disinfection system is
available or if the water has to be used without the agent for
disinfecting the water paths. See "Microbiological water test" 
[→ 216].