4.15.1.4

Switching the key sound on/off

A setting can be made to issue an acoustic signal when the operator
presses a button on the dentist or assistant element.
ü The setting focus Key tone flashes.
1. Change the value with the favorite keys 1 or 3.
Ä If there is a 1 in the setting focus, the key tone is switched on.
2. Confirm with the 2 / Setup favorite key.
4.15.1.5

Coupling suction to the 4-way foot switch.

If the treatment center is equipped with a position selector valve for the
suction system, it can be set so suction flow can be interrupted or
restarted by moving the 4-way foot control switch at the base of the
chair in any direction. Follow the safety instructions, see "Suction
handpieces"  [→ 152]
ü The setting focus Couple suction at 4-way foot switch flashes.
1. Change the value with the favorite keys 1 or 3.
Ä If a 1 is in setting focus, the suction can be turned on/off at the
4-way foot switch.
2. Confirm with the 2 / Setup favorite key.
If you put the suction hose in the holder while the suction flow is
interrupted, the suction flow is automatically restarted when you pick it
up again.
4.15.1.6

Switching the water heater on/off

The optional heater for the treatment water can be switched on/off.
ü The water heater setting focus flashes.
1. Change the value with the favorite keys 1 or 3.
Ä If there is a 1 in the setting focus, the water heater is switched on.
2. Confirm with the 2 / Setup favorite key.
4.15.1.7

Setting the purging time for the purge function

The purging time for the purge function may be set between 20 and 120
seconds.
ü The setting focus purge function flashes.
1. Use favorite keys 1 and 3 to set the purge time.
2. Confirm with the 2 / Setup favorite key.
4.15.1.8

Setting the purging time for the autopurge function

The autopurge function is only available on the Intego Pro treatment
center.
The purging time for the autopurge function can be set between 60 and
180 seconds.
ü The setting focus purge function flashes.
1. Use favorite keys 1 and 3 to set the purge time.
2. Confirm with the 2 / Setup favorite key.