5 Care, cleaning and maintenance by the practice team
5.1 Basics
218
Manual drying
No drying is required following wiping with disinfectants since surplus
disinfectant evaporates. Surplus water from the cleaning process can
be removed with a soft cloth.
Sterilization
Sterilization may be conducted for components that are marked
accordingly. Steam sterilizers that fulfill the requirements of EN 13060,
class B (e.g., DAC Premium / DAC Professional) are approved.
The sterilization must be completed with multiple vacuum fractionation
(class B sterilizer). The process parameters can be found in the
engraved characters on the relevant components and the instructions
for use for these.
During the drying cycle, the sterilized parts must not exceed a
temperature of 140 °C (284 °F) during the drying.
5.1.5

Inspection, maintenance and testing

Unless otherwise specified in this operating manual, test all components
for proper functioning on a regular basis and carry out a visual
inspection for damage and wear. Exchange damaged components if
necessary.
Dentsply Sirona
Operating Instructions Intego
64  5 7  4 98 D3543
D3543.201.01.08.02    02.2020