2 Safety instructions

2.3 On-site installation

16
On-site installation
2.3
The on-site installation must have been performed according to our
requirements. The details are described in the document "Installation
Requirements".
2.4

Installation of the treatment center

Installation must be carried out by authorized personnel according to
the installation instructions.
2.5

Media quality

The air and water supplies must meet the requirements specified in the
installation requirements.
As the operator of the treatment center, you are generally responsible
for the water quality.
The bacterial count must comply with the national regulations for
drinking water and must not exceed 500 CFU/ml under any
circumstances (CFU: colony forming unit).
If the bacterial count is too high, the building water system must be
checked and the cause of contamination eliminated. It may be
necessary to install an external system for self-sufficient water supply or
reprocessing of the process water. Alternatively, for Intego the optional
fresh water bottle, or for Intego Pro the empty disinfectant tank can be
used as a water container for a self-sufficient water supply. For
treatment centers with the Ambidextrous option, depending on the
equipment, the fresh water bottle or the empty disinfectant tank can be
used as a water container.
Before installation of the treatment center, an acceptable
microbiological water quality for the domestic water supply must be
ensured and documented in the form of the bacterial count. Sampling
and germ counts should be performed by a competent laboratory.
Test the water quality from the treatment center at regular intervals and
after it has not been used for >1 week, see "Microbiological water test" 
[→ 216]. Please contact your specialized dealer or your dental
association for the respective national requirements and measures.
Further information on the media quality requirements are listed in the
"Installation requirements" document.
Highly immunosuppressed patients or patients with specific pulmonary
diseases should not come in contact with the water of the treatment
unit. The use of sterile solutions is recommended.
2.6

Connection to the public drinking water system

Treatment center with isolation from public drinking water supply
The Intego Pro treatment center, provided it is equipped with a
disinfection system, complies with the requirements of EN 1717 (free
discharge with an isolation distance ≥ 20 mm) and the DVGW German
Gas and Water Association). It is intrinsically safe in accordance with
worksheet W540 and therefore also fulfills the W270 requirements as