5.6.2

Sanitizing with disinfection system

Sanitizing, that is the disinfection of the treatment water paths can be
performed by using the disinfection system. During this sanitation, the
treatment water shall first be emptied according to a pre-defined
process, the agent for disinfecting the water paths is then filled
undiluted into the treatment water paths and this is then flushed out
again at the end. Sanitizing takes at least 24 hours and should not
exceed 3 days.
Sanitizing must be performed:
● Regularly every 4 weeks when operating with the public drinking
water supply or if stand-alone water supply is used more than 28
days in exceptional cases
If the treatment center features a disinfection system, the display
Days until next sanitation appears on the EasyPad or touchscreen. It
first appears three days before the date of sanitation.
● After longer periods of disuse (> one week)
● If the bacterial count exceeds 100 bacteria per milliliter, see
"Microbiological water test"  [→ 216].
● After changing from stand-alone water supply to public drinking
water supply and the agent for disinfecting the water paths
NOTE
Additional devices connected to the external device connection must
not be sanitized with the treatment center.
The additional devices could be damaged. Residues of the agent for
disinfecting the water paths can remain in the additional devices.
➢ Disconnect any additional devices from the treatment center prior
to sanitization.
Preparation
The following preparations should be made prior to beginning
sanitation.
1. If your treatment center is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water
paths.
2. Set all burr drives and the scaler to the maximum water flow rate.
The water regulators under the instrument holder must be turned
counterclockwise.
IMPORTANT
The water flow rate to the instruments is not checked by the treatment
center.
➢ Make sure that the maximum water flow rate is set for all
instruments to be purged.
3. Put all instruments and suction hoses in place.
4. Do not remove the tumbler holder from the cuspidor. Put an empty
glass with a volume ≥200 ml under the tumbler outlet to prevent
discoloration from the agent for disinfecting the water paths.
5 Care, cleaning and maintenance by the practice team
5.6 Sanitizing