On-site installation ....................Substrate, floor..................... 3.1.1 Load-bearing capacity of the floor..........3.1.2 Stability ..................3.1.2.1 With Intego and Intego pro ........... 3.1.2.2 With Intego Ambidextrous and Intego pro Ambidextrous..............Requirements of the supply media .............. 3.2.1 Requirements of the water supply ..........
Page 3
Distances in the Intego pro treatment room with Heliodent Plus unit model ..................4.2.7 Side view of Intego pro with Heliodent Plus unit model ....Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 4.3.1 Distances in the Intego Ambidextrous, Intego pro Ambidextrous treatment room................
General information Notes on the installation prerequisites This document describes the installation prerequisites for all versions of the Intego treatment center. It contains the following information: ● Required information for practice planning. ● Information for the installer and operator regarding the necessary quality of the air and water supply media.
1 General information Dentsply Sirona 1.2 Structure of the document Installation requirements Intego Structure of the document 1.2.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
For reasons of product safety, this product may be operated only with original Dentsply Sirona parts or third-party parts expressly approved by Dentsply Sirona. In the event of changes which were not foreseen, Dentsply Sirona is not liable for resulting damages.
3 On-site installation Dentsply Sirona 3.1 Substrate, floor Installation requirements Intego On-site installation Substrate, floor Unevenness The floor must be level and horizontal in accordance with DIN 18 202. The compensating plates (A) in the accessory pack can be used for slight unevenness in the floor.
3.1 Substrate, floor 3.1.2 Stability 3.1.2.1 With Intego and Intego pro With a concrete floor ● Use a 12mm dia. masonry drill to drill the holes. ● The depth of the holes depends on the condition of the floor. For concrete floor without screed...
3.1 Substrate, floor Installation requirements Intego 3.1.2.2 With Intego Ambidextrous and Intego pro Ambidextrous With a concrete floor ● Use a 12mm dia. masonry drill to drill the holes. ● The depth of the holes depends on the condition of the floor.
Dentsply Sirona 3 On-site installation Installation requirements Intego 3.1 Substrate, floor With a wooden floor ● Use a 7 mm dia. wood drill to drill the holes. ● Hole depth is: For wooden floor (D): Hole depth 120 mm (G) 6 or 8 screws 110 mm long (E)
3 On-site installation Dentsply Sirona 3.2 Requirements of the supply media Installation requirements Intego Requirements of the supply media * = For use with an Air Venturi air jet pump Suction machine Compressed air (free from oil) The compressor must be able to draw in air in an efficient, hygienic manner.
With In- tego, the optional fresh water bottle can be used as a water container, as can the empty disinfec- tant tank with Intego Pro. Sampling and the colony count must be performed by a competent laboratory.
Connection to the public drinking water system Treatment center isolated from the public drinking water supply The Intego Pro treatment center, if it is equipped with a disinfection system, complies with the requirements of EN 1717 (free discharge with an isolation distance ≥ 20 mm) and the DVGW German Gas and Water Association).
Dentsply Sirona 3 On-site installation Installation requirements Intego 3.2 Requirements of the supply media 3.2.1.3 Installation of the water supply on the building side ● Carry out the installation in accordance with the national installation requirements (e.g. EN 1717). ● Connect the treatment center to the cold water supply.
3 On-site installation Dentsply Sirona 3.2 Requirements of the supply media Installation requirements Intego 3.2.2 Requirements of the air supply The air supply is required to drive the turbines, cool the drill drives, for the cooling spray and to control the treatment center.
– 11 mm: Spray aspirator ● INTEGO: Filter mesh size 1.5 x 3.0 mm ● INTEGO pro: Filter mesh size 1.3 x 1.3 mm ● Sirona recommends using cannulae with a bypass in order to prevent return flow in adhered cannulas.
3 On-site installation Dentsply Sirona 3.2 Requirements of the supply media Installation requirements Intego 3.2.3.1 Typical pressure loss in the suction system Suction flow of large suction hose red (○): Only large suction device black (◊): Saliva ejector in parallel 3.2.4...
Dentsply Sirona 3 On-site installation Installation requirements Intego 3.3 Above-floor installation of supply lines Above-floor installation of supply lines The supply lines can be installed above the floor (above-floor installation) or through the floor (underfloor installation). For the installation of the supply lines through the floor, see "Underfloor installation of supply lines" ...
3 On-site installation Dentsply Sirona 3.3 Above-floor installation of supply lines Installation requirements Intego In the case of re-ordering for a retrofit ● For connection from the wall, the above-floor installation kit and a diaphragm are required: – Above-floor installation kit (D) Order No.: 64 43 746 and –...
3 On-site installation Dentsply Sirona 3.4 Underfloor installation of supply lines Installation requirements Intego Underfloor installation of supply lines 3.4.1 Installation template We recommend that you order the installation template (REF 33 15 830) from Sirona for laying the pipe ends in the installation field.
3 On-site installation Dentsply Sirona 3.4 Underfloor installation of supply lines Installation requirements Intego 3.4.2 Installation of the supply lines in the termination panel ü An installation template is available or was prepared. 1. Check the position of the supply lines against the installation template as per the practice blueprint.
Dentsply Sirona 3 On-site installation Installation requirements Intego 3.5 Cleaning the air and water pipes Cleaning the air and water pipes NOTE Chips and other foreign materials could be flushed and/or blown into the treatment center. Metal chips can cause malfunctions of the pneumatic components.
3 On-site installation Dentsply Sirona 3.6 Potential equalization Installation requirements Intego Potential equalization The treatment center does not require potential equalization The treatment center, monitor, PC and devices connected to the external device connection have the same potential as a result of the wiring of the treatment center.
Dentsply Sirona 3 On-site installation Installation requirements Intego 3.7 Underfloor installation of the PC connections Underfloor installation of the PC connections Depending on the prevailing local conditions, the existing cable set can be installed in the cable duct of an underfloor installation by an installer.
3 On-site installation Dentsply Sirona 3.7 Underfloor installation of the PC connections Installation requirements Intego ➢ Pull the USB , Ethernet, HDMI cables and protective ground wire from the treatment center through the cable duct to the location of the PCs C.
4 Dimensions, technical data Dentsply Sirona 4.1 Intego dimensions, scale 1:20 Installation requirements Intego 4.1.2 Intego side view IMPORTANT Recommended room height ≥ 2220 mm Dimensions at the Compact water unit Monitor on lamp support tube Height of dentist element TS or CS...
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.1 Intego dimensions, scale 1:20 4.1.6 Distances in the Intego treatment room with Heliodent Plus unit model Recommended distances from cabinet or wall. Center of the floor cut-out/installation area IMPORTANT! The lamp installed here and the dentist element with/without a tray and the Heliodent Plus have a swivel range which exceeds the specified distances.
Side view of Intego with Heliodent Plus unit model Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for radiation cable The radiation cable between Intego and the wall adapter is supplied (12.5 mm outer diameter, max. cable length 10 me- ters (393”). CAUTION The Heliodent Plus radiation cable must not be laid together with USB or HDMI cables.
4 Dimensions, technical data Dentsply Sirona 4.2 Intego pro dimensions, scale 1:20 Installation requirements Intego 4.2.2 Intego pro side view IMPORTANT Recommended room height ≥ 2220 mm Dimensions at the Comfort water unit (height and swivel range) Monitor on lamp support tube...
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.2 Intego pro dimensions, scale 1:20 4.2.5 Top view of Intego pro with options TS dentist element CS dentist element 64 57 092 D3543 D3543.021.01.09.02 12.2018...
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.2 Intego pro dimensions, scale 1:20 4.2.6 Distances in the Intego pro treatment room with Heliodent Plus unit model Recommended distances from cabinet or wall. Center of the floor cut-out/installation area IMPORTANT!
Dentsply Sirona 4.2 Intego pro dimensions, scale 1:20 Installation requirements Intego 4.2.7 Side view of Intego pro with Heliodent Plus unit model Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for radiation cable The radiation cable between Intego and the wall adapter is supplied (12.5 mm outer diameter, max.
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 4.3.1 Distances in the Intego Ambidextrous, Intego pro Ambidextrous treatment room Recommended distances from cabinet or wall.
Page 46
4 Dimensions, technical data Dentsply Sirona 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 Installation requirements Intego Dentist element TS or CS IMPORTANT! The minimum distance of the rear edge of the chair (with cladding) to the wall or cabinets must be complied with, other- wise the water unit cannot be fully swiveled round.
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 4.3.2 Side view of Intego Ambidextrous, Intego pro Ambidextrous IMPORTANT Recommended room height ≥ 2220 mm Dimensions at the Compact water unit...
4 Dimensions, technical data Dentsply Sirona 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 Installation requirements Intego 4.3.3 Top view of Intego Ambidextrous, Intego pro Ambidextrous with dentist element CS Head support: Length of treatment center with patient height of 176 ...
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 4.3.4 Top view of Intego Ambidextrous, Intego pro Ambidextrous with dentist element TS Head support: Length of treatment center with patient height of 176 ...
4 Dimensions, technical data Dentsply Sirona 4.3 Dimensions for Intego Ambidextrous, Intego pro Ambidextrous, scale 1:20 Installation requirements Intego 4.3.5 Top view of Intego Ambidextrous, Intego pro Ambidextrous with options TS dentist element CS dentist element...
Adapter plate Adapter plates are available for replacement of one of the following dental treatment centers for a right-handed person with a Intego / Intego Pro treatment center: ● C2, C3, C4, C6, C8 (in the case of C4, year of manufacture 1999-2001, a new drill hole must be drilled in the area of the water unit (C) .)
D on the floor using two screws. Demo chair plate thickness: 10 mm REF 64 46 061 Intego / Intego Pro demo chair plate INTEGO Ambidextrous / INTEGO pro Ambidextrous demo chair plate For floors which do not permit permanent connection of the unit (e.g.
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.5 Information on planning for the practice Information on planning for the practice The following file(s) for practice planning are available for download in the dealer area of the Sirona website under DOWNLOADS => CAAD files: ●...
4 Dimensions, technical data Dentsply Sirona 4.6 Technical data Installation requirements Intego Technical data Model designation: Intego / Intego Pro Power connection: 100 – 240 V AC ± 10% 50/60 Hz Rated current: 3.3 A – 1.5 A at 100 – 240 V also max. 6 A for external devices...
Dentsply Sirona 4 Dimensions, technical data Installation requirements Intego 4.6 Technical data 4.6.1 Operating, transport and storage conditions Mode of operation Continuous operation with intermittent loading corresponding to the dental mode of working. Drive motors for chair operation: intermittent use, max. 2 minutes on and 18 minutes off Permanently connected unit.
Dentsply Sirona 4.6 Technical data Installation requirements Intego 4.6.2 Weight and packaging Dimensions of the packaging Weight (with packaging and ac- cessories/without packaging): Intego CS dentist element: 120 cm x 80 cm x 92 cm 37 kg / 24.5 kg Intego TS dentist element: 120 cm x 80 cm x 71 cm 36 kg / 24 kg Comfort assistant element: 87 cm x 31 cm x 82 cm 10.4 kg / 6 kg...
“Connection to the public drinking water supply” [→ 14] Original language of this document: German ANVISA registration number: 80745400024 Intego / Intego Pro bear the CE marking in accordance with the provisions of Council Directive 93/42/EEC of June 14, 1993 concerning medical devices.
4 Dimensions, technical data Dentsply Sirona 4.7 Standards/Approvals Installation requirements Intego When equipped with a disinfection system or fresh water bottle without switchover to the public drinking water supply, the treatment center complies with the technical rules and requirements on safety and hygiene for connection to the public drinking water supply.
5.1 Accessories Electromagnetic compatibility Observance of the following information is necessary to ensure safe operation regarding EMC aspects. Intego complies with the requirements for electromagnetic compatibility (EMC) according to IEC 60601-1-2:2014 Intego is hereinafter referred to as "UNIT". Accessories Connecting the PC The required interface cables can be ordered from Sirona.
5 Electromagnetic compatibility Dentsply Sirona 5.2 Electromagnetic emission Installation requirements Intego Accessories for EMC measurement The EMC measurements were performed with the following PC: PC as peripheral device for checking Fujitsu the interfaces with: ESPRIMO P710 E90+ PC equipment: Processor Intel Core i5 3470;...
Dentsply Sirona 5 Electromagnetic compatibility Installation requirements Intego 5.3 Immunity to interference Immunity to interference The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
5 Electromagnetic compatibility Dentsply Sirona 5.4 Working clearances Installation requirements Intego Working clearances The UNIT is intended for operation in an electromagnetic environment, where radiated RF interference is checked. The customer or the user of the UNIT can help prevent electromagnetic interference by maintaining...
We recommend performing an inspection of the circumstances on location at least 4 weeks prior to installation. The checklist should help you when doing this. This can help ensure a smooth procedure on the day that the Intego / Intego Pro is actually installed. Installation location: ●...
● Compensating plates can be used for slight unevenness, no mounting plates required. ☐ ● Adapter plates: Intego Pro REF 64 32 061, Intego REF 65 42 018: Can be used to replace ☐ certain C-line dental treatment centers, for details see Mounting plates [→ 52] ●...
Dentsply Sirona 6 Checklist Installation requirements Intego 6.3 IT hardware IT hardware Mode of operation ☑ ● PC in treatment room ☐ PC system requirements: ☑ IT hardware and software requirements are described in the document "Installation notes and ☐...
☐ ● Network connection for external PC available. ☐ IMPORTANT! The use of routers between Intego / Intego Pro and the treatment center PC must be avoided. ● Network configuration plan available. ☐ ● Network jacks have been certified. ☐...
● Is there a DHCP server (dynamic TCP/IP address assignment)? ☐ Yes ☐ No IMPORTANT! A static address should be assigned to Intego / Intego Pro! It must not lie in the dynamic address range! ● Intego / Intego Pro: _____ . _____ . _____ . ____ ●...