5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
246
ü The housing, keyboard and nozzle are cleaned and disinfected.
➢ Sterilize the housing, keyboard and nozzle in the steam sterilizer
with saturated water vapor.
Overpressure:
Temperature:
Holding time:
Steam sterilizers that fulfill the requirements of either EN 13060, class B
(for example, DAC Premium / DAC Professional) or EN 13060, class S
and are suitable for the sterilization of 3-way syringes are approved.
Steam sterilizers that fulfill the requirements of either EN 13060, class B
(for example, DAC Professional / DAC Universal) or EN 13060, class S
and are suitable for the sterilization of 3-way syringes are approved.
NOTE
Do not exceed 140°C (284°F), even during the drying phase.
Please observe the instructions for use for the sterilizer.
After sterilizing:
1. Remove the housing, keyboard and nozzle from the steam sterilizer
immediately.
CAUTION
These parts are hot. Risk of burns!
NOTE
Do not try to accelerate the cool-down process by placing the parts in
cold water. This will damage the parts!
2. Store all components so that they are protected against
contamination.
3. Sterilize again once the storage period has elapsed.
5.3.5.4

Disinfecting/sterilizing motors and adapters

NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents that have been
approved by Dentsply Sirona, see "Care, cleaning, and disinfecting
agents"  [→ 216]!
Use disinfectants and other agents that contain no corrosive
components such as chloride.
All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Use only disinfectants
with no protein-fixing properties.
NOTE
Never clean in an ultrasonic bath!
Never immerse in disinfectants!
2.04 bar (29.59 psi)
134°C (274°F)
3 min
D3543.201.01.08.02    02.2020
Dentsply Sirona
Operating Instructions Intego
64  5 7  4 98 D3543