3 Unit description

3.2 Technical data

24
Technical data
3.2
Model designation:
Power connection:
Rated current:
Type of ground connection:
Overvoltage category:
Average power consump-
tion (for dimensioning an air
conditioning system):
Power consumption when
switched off:
Main building fuse:
Protection class:
Device class in accordance
with Directive 93/42/EEC:
Degree of protection
against electrical shock:
Degree of protection
against ingress of water:
Mode of operation
Operating Instructions Intego
Intego / Intego Pro
100 – 240 V AC ± 10%
50/60 Hz
3.3 A – 1.5 A at 100 – 240 V
also max. 6 A for external devices
TN-C-S system or TN-S system
(acc. to IEC 60364-1)
2 acc. to IEC 60664-1
100 W
0 W (power switch present)
Type B automatic circuit breaker
100 - 115 VAC: 20 A medium-blow
220 – 240 V AC: 16 A medium-blow
Class Idevice
Class IIa equipment
Type B applied parts
External intraoral camera SiroCam F /
AF / AF+. These are:
Applied part type BF
Ordinary equipment (without protection
against ingress of water)
The foot switch has an IPX1 degree of
protection against liquids (drip-proof).
Continuous operation with intermittent
loading corresponding to the dental
mode of working.  [→ 18]
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.
D3543.201.01.08.02    02.2020