2 Safety instructions

2.17 Modifications and extensions of the system

20
Modifications and extensions of the system
2.17
For reasons of product safety, this product may be operated only with
original Dentsply Sirona parts or third-party parts expressly approved by
Dentsply Sirona. In the event of changes which were not foreseen,
Dentsply Sirona is not liable for resulting damages.
All units connected to this product must comply with the applicable
standards:
● IEC 60601-1, Medical electrical equipment
● IEC 60950-1, Information technology equipment
● IEC 62368-1, Audio/video, information and communication
technology equipment

Electromagnetic compatibility

2.18
Medical electrical devices are subject to special precautionary
measures with regard to electromagnetic compatibility (EMC). They
must be installed and operated as specified in the document
"Installation Requirements".
Portable HF communication equipment, including accessories, should
not be used in the vicinity of the device. Non-observance can impair the
performance of the device.
Operating an HF surgical device
Treatment with HF surgical devices creates strong electromagnetic
fields, which may affect other electronic devices. Do not place external
HF surgical device on the surfaces of the treatment center. Do not run
the cable of the HF handpiece over the surfaces of dentist or assistant
element, user interfaces, or other keys such as the On/Off key on the X-
ray image viewer. Electromagnetic interference can often be reduced by
operating the external HF surgical device with a neutral electrode.
Turn on the treatment center with the Clean button inactive, to avoid
malfunction. See section "Display mode / Clean"  [→ 41] on the
EasyPad or "Clean button"  [→ 44] on the EasyTouch.
Sivision Digital and USB interface
The presence of electromagnetic interference in the vicinity of the
treatment center may cause image degradation and interruptions in the
data transmission via the USB interface to the PC. In such cases,
repeat the image recording or other operations.
In the event of strong interference, it may be necessary to restart the
PC and treatment center. Do not use the PC for controlling other
devices that perform essential tasks.