Trouble-Free Operation; Vacuum System; Patient Chair; Intermittent Operation - Dentsply Sirona Intego Operating Instructions Manual

Hide thumbs Also See for Intego:
Table of Contents

Advertisement

2 Safety instructions

2.8 Trouble-free operation

max. load of chair
max. load of chair
140 kg
185 kg

Intermittent operation

Intermittent operation
of chair motors:
of chair motors:
2 min ON/18 min OFF
2 min ON/18 min OFF
ü
max. load of chair
max. load of chair
140 kg
185 kg
Intermittent operation
Intermittent operation
of chair motors:
of chair motors:
2 min ON/18 min OFF
2 min ON/18 min OFF
18
Trouble-free operation
2.8
This unit may only be used in proper working condition. If trouble-free
operation cannot be ensured, e.g., due to malfunctions, noticeable or
different noises, or damage, the unit must be taken out of service,
checked by authorized technicians for malfunctions and, if necessary,
repaired or replaced.

Vacuum system

2.9
The suction removal of aluminum and other metal oxides from blasting
devices via the amalgam separator and the automatic separator
installed in the treatment center is prohibited! This would result in
extreme wear and clogging of the vacuum and drain lines.
A separate vacuum system must be used in connection with metal
oxide blasting devices. Treatment centers equipped with a central
standard wet suction are generally suitable for suction removal of the
above material. However, make sure to observe the instructions
provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in connection with
Dentsply Sirona treatment centers. However, in such cases, make sure
that the system is subsequently flushed with an adequate amount of
water.

Patient chair

2.10
Please observe the maximum load capacity for the patient chair.
140 kg (308.6 lbs) forIntego, 185 kg (407.9 lbs) as option
185 kg (407.9 lbs) forIntego Pro
The permissible maximum load capacity is indicated on a plate next to
the rating plate of the treatment center.
The weight distribution complies with ISO 6875. The mechanical
stability is tested with a multiple safety factor acc. to IEC 60601-1.
The maximum permissible weight of accessories mounted on the
patient chair is 5 kg (11 lbs).
The patient's arms and legs must be resting on the upholstery of the
û
chair.
Intermittent operation
2.11
The motors of the treatment center and of the treatment instruments are
designed for intermittent operation corresponding to the dental mode of
treatment.
Drive motors for patient chair and backrest: max. 10% duty cycle, (max.
2 min "ON" / 18 min "OFF")
Dentsply Sirona
Operating Instructions Intego
64  5 7  4 98 D3543
D3543.201.01.08.02    02.2020

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intego proIntego ambidextrousIntego pro ambidextrous

Table of Contents