To remove the probe adjuster assembly probe and source enclosure seals
1.
Remove the source enclosure from the instrument (see
source enclosure from the instrument" on page
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly:
•
Probe seal
•
Nebulizer gas seal
See
"Removing O-rings and seals" on page
3.
Use the O-ring removal kit to carefully remove the following seals from
the source enclosure:
•
Source enclosure seal
•
Nebulizer gas seal
4-76
Maintenance Procedures
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Probe adjuster nebulizer gas seal
4-8).
4-14.
Probe adjuster assembly probe seal
"Removing the