Waters Xevo TQ MS Operator And Maintenance Manual page 113

Table of Contents

Advertisement

5.
Use the 10-mm wrench to remove the probe tip.
If the probe tip is difficult to remove, use the 7-mm wrench in
Tip:
conjunction with the 10-mm wrench.
Probe tip
7-mm wrench
Replacing the ESI probe sample capillary
10-mm wrench
10-mm wrench
Probe tip
4-65