Waters Xevo TQ MS Operator And Maintenance Manual page 116

Table of Contents

Advertisement

PTFE liner tube
Conductive sleeve
Red, PEEK tubing
Sharp knife
To install the new capillary
Warning:
handle the probe with care.
1.
Use the sharp knife to cut an approximately 60 cm (24 inches) piece of
red, PEEK tubing squarely (that is, perpendicular to its horizontal axis).
2.
Insert one end of the red, PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
4-68
Maintenance Procedures
The probe and source components can be contaminated
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The ESI probe tip is sharp. To avoid puncture wounds,
Doing so ensures a minimum dead volume when fitting the
Probe inlet connector
PEEK tubing
TP02671