3
Troubleshooting and Test Functions
Encoder Missing
The optical encoder on the pump motor in the quaternary pump is missing or
defective.
The processor checks the presence of the pump encoder connector every
2 seconds. If the connector is not detected by the processor, the error message
is generated.
Probable Causes
• Defective or disconnected pump encoder connector.
• Defective pump drive assembly.
Suggested Actions
70
Ensure the connector is clean, and seated correctly.
Exchange the pump drive assembly.
1200 Series QP Reference Manual