Repairing the Pump

Assembling the Main Cover

When required
Tools required
Parts required
The cover kit contains all parts, but it is not assembled.
N O T E
In case you insert the left or right side in the opposite position, you may not be able
WA R N I N G
to remove the side from the top part.
1 Place the top part on the bench and insert the left and right
side into the top part.
3 Replace the quaternary pump into the stack and reconnect the cables and capillaries.
4 Turn on the quaternary pump.
156
If cover is broken
None
Cover kit 5065-9985 (includes base, top, left and right)
2 Replace the cover.